<strong>Cardell</strong>® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualSECTION 8RESPIRATION MONITORING8.1 General InformationThe monitor provides two respiration monitoring methods: thoracicimpedance (indirect) and through the CO2 microstream sampline line(direct).NOTE: If the patient is intubated, direct respiration monitoring through theCO2 sample line is recommended. If you choose to monitor respirationusing the thoracic impedance method, place the ECG electrodes on thepatient’s trunk for more reliable readings.8.2 Respiration SetupTo enter respiration setup menu,Fig. 8-1 Respiration Setup MenuSetup →Rotate the knob→ RESP → Press the knobRespiration setup menu is as figure on the above.RESP Lead : While ECG is in Three Lead mode, the lead selections availableare RA-LA/RA-LL; if ECG is in Five Lead mode, RA-LA/RA-LL/RL-LA/62 M1.0 5-15-09