CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s Manual7.6.2 Parameter Adjustment Range:ParameterHR LimitAdjustment Range15-300 bpm7.8.3 Abnormal Status Alarm:Abnormal Status alarm includes “Asystole” and “lead off”.CAUTION: When ECG amplitude is too low, it may result in inaccurateheart rate or Pseudo asystole. We suggest when ECG waveform is too low,change the lead, and adjust to the ECG lead which has maximum amplitude.Otherwise, the monitor may give “Asystole” alarm.7.9 PrecautionsWARNING: When using defibrillator, make sure the electrode and patientcable are not in contact with metal or other conductor surface or groundingdevices. During defibrillation, do not touch patient, table or instrument.WARNING: Ensure conductive parts including electrodes of the patientcable do not come into contact with any conductive surfaces.WARNING: Do not use the patient monitor during MRI or CT scan.CAUTION: Leads and cables should be away from patient’s neck.M1.0 5-15-0959