CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualNOTE: When using three leads, it only detects HR based on ECG1 signals.Lead ModeThree lead and Five lead are options (standard configuration is three lead).When selecting Five Lead, the ECG leads you can choose are: Ⅰ/Ⅱ/Ⅲ/V/aVR/aVL/aVF and the RESP leads you can choose are: RA-LA, RA-LL,LA –RL, LL- RL.When selecting Three Lead, the ECG leads you can choose are: Ⅰ/Ⅱ/Ⅲ andthe RESP leads you can choose are: RA-LA, RA-LL.Work ModeMonitoring /Surgical /Diagnosis mode is for the needs of differentcircumstances. User may set in accordance with the actual needs.Diagnosis mode:To display the original ECG waveform which is not filtered.Monitoring mode:To display the waveform where interference has been filtered.Surgical mode:This is only for the circumstances where there are significant interferencesoutside and ECG waveform has significant distortion (for example operatingroom). It is not calibration significant.Alarm sound On/OffM1.0 5-15-0957