CARDELL® <strong>MAX</strong>-<strong>12</strong><strong>HD</strong> User’s ManualThe above setup will remain after power off.13.4 Transducer Connection1. When the device is turned on, the IBP channel will display “No transducer”and alarm sound can be heard.2. Plug transducer cable into the IBP or IBP2 socket, and alarm will stopindicating the connection is correct. The other end of the transducer cable isconnected as follows:1 2 3 4 5Fig. 13-4 IBP Transducer Connection Diagram1 The T (1) is used to make (2) to be open to the air.2 The T (3) is used to block (2) from (3) and (4).3 The pressure monitoring tube (4) is to ensure the accuracy of themeasurement.4 The (5) in the above diagram is to connect patient catheter.1. Fill in the catheter system from T (3) and make sure there is no bubble inthe system.2. Connect patient catheter to pressure monitoring tube, make sure there isno air in catheter, pressure monitoring tube or transducer.M1.0 5-15-09103