43performed at Inframetrics to determine the step response <strong>of</strong> the TEC pump. Thetemperature sensing diodes were connected as shown in Figure 4.6. The two diodes wereconnected in series and biased with 102K0 resistor from +5 volts with current <strong>of</strong> about40p.A. The output <strong>of</strong> the diodes were connected to an oscilloscope to measure the stepresponse <strong>of</strong> the TEC pump due to heating and cooling <strong>of</strong> the thermoelectric pump.Figure 4.6 The setup for determining the diode response.The thermoelectric cooler was connected to a power supply, and the step inputwas simulated by turning on the power supply. The oscilloscope was set to trigger on therising edge <strong>of</strong> the input pulse to record the response data. Figure 4.7 shows the resultingresponse curves <strong>of</strong> the TEC pump, where the vertical scale represents the output voltage<strong>of</strong> the temperature sensors (diodes) and the horizontal scale represents time. In Figure4.7 a), the TEC was cooled without a heat sink.