Analysis of 320X240 uncooled microbolometer focal plane array ...
Analysis of 320X240 uncooled microbolometer focal plane array ...
Analysis of 320X240 uncooled microbolometer focal plane array ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
43performed at Inframetrics to determine the step response <strong>of</strong> the TEC pump. Thetemperature sensing diodes were connected as shown in Figure 4.6. The two diodes wereconnected in series and biased with 102K0 resistor from +5 volts with current <strong>of</strong> about40p.A. The output <strong>of</strong> the diodes were connected to an oscilloscope to measure the stepresponse <strong>of</strong> the TEC pump due to heating and cooling <strong>of</strong> the thermoelectric pump.Figure 4.6 The setup for determining the diode response.The thermoelectric cooler was connected to a power supply, and the step inputwas simulated by turning on the power supply. The oscilloscope was set to trigger on therising edge <strong>of</strong> the input pulse to record the response data. Figure 4.7 shows the resultingresponse curves <strong>of</strong> the TEC pump, where the vertical scale represents the output voltage<strong>of</strong> the temperature sensors (diodes) and the horizontal scale represents time. In Figure4.7 a), the TEC was cooled without a heat sink.