<strong>Product</strong> <strong>Manual</strong> <strong>for</strong> <strong>IonPac</strong> <strong>AS23</strong> amd <strong>IonPac</strong> AG23 <strong>Columns</strong> Page 28 of 32For DX-300 systems equipped with a Rheodyne Microinjection Valve, Model 9126 (<strong>Dionex</strong> P/N 044697), consult theaccompanying manual <strong>for</strong> service instructions.6.3.5 Poor Resolution of Only Phosphate and SulfateA. Causes1. Sodium carbonate is contaminated with sodium hydroxide,2. Inadequate equilibration after use of an alkaline buffer or hydroxide eluent,3. Sodium carbonate was dried at temperatures > 110°C.B. Action1. Use <strong>Dionex</strong> <strong>AS23</strong> Eluent Concentrate (P/N 064161).2. Use a high purity sodium carbonate salt.3. Dry the sodium carbonate at a lower temperature. See section 4.3.1 and section 5.1.Document No. 065120-05 ©2012 Thermo Fisher Scientific July 2012