Measured value (ng/g)Evaluation of immunoassay kits for determination of ochratoxin A in cereals10090807060504030201000 20 40 60 80 100 120Reference value (ng/g)Figure 5: Linear fit of measured OTA concentration versus reference value for CharmTable 19: Summary of the ordinary linear regression for CharmCoefficientsSlope 0.783Intercept 1.837R 2 0.988ρc 0.960T-test 0.7503.1.4.5.2.1.3 ConclusionAn overall good agreement between the reference values is obtained, only at low or high concentrations there issome deviation.3.1.4.5.2.2 R-Biopharm3.1.4.5.2.2.1 Descriptive comparisonThe results (Figure 6) shows random scattering around the ordinate.CODA-CERVA | Administrative evaluation 3-23