GUIDO RAYOS X, S.A.40Turn on the Radiant Warmer by pressing the I/O Switch located at the rear side of thetrolley.Press the ON/OFF key.Press three times the "PRE." key.The LCD will display :AT=00.0 PT=00.0OX=00.0 RH=00.0(AT: Air Temp., PT: Patient Temp., OX: Oxygen, RH: Non operative parameter)A. Air Temperature• Place a jumper at J15.• Set Simulator switch at "0".• Set DVM probes between GND-TP8 and TP1. Twist P1 Potentiometer to readbetween 2 to 4 mV.• Set Simulator switch at "40/99".• Twist P2 Potentiometer to read 40.0 ºC at the Air Temperature Display on theIncubator Control.• Set Simulator switch at "18/21".• Twist P1 Potentiometer to read 18.0 ºC at the Air Temperature Display on theIncubator Control.Because adjustments are mutual interactive, therefore have to be repeated steps 3 and 6until to get the correct value.B. Patient Temperature• Set Simulator switch at "0".• Set DVM probes between GND-TP8 and TP2. Twist P5 Potentiometer to readbetween 2 to 4 mV.• Set Simulator switch at "40/99".• Twist P4 Potentiometer to read 40.0 ºC at the Air Temperature Display on theIncubator Control.• Set Simulator switch at "18/21".• Twist P5 Potentiometer to read 18.0 ºC at the Air Temperature Display on theIncubator Control.Because adjustments are mutual interactive, therefore have to be repeated steps 3 and 6until to get the correct value.C. Oxygen Concentration• Set Simulator switch at "0".• Set DVM probes between GND-TP8 and TP5. Twist P9 Potentiometer to read5 mV maximum.MAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!