GUIDO RAYOS X, S.A.26the PEEP valve.Adjust the PEEP cap to set the desired PEEP level as displayed on themanometer.8 Remove the tst lung from the T-piece. Fit the desired neonatal resuscitation maskor endotracheal tube to the T-piece. The Neopuff resuscitator is now ready foruse.9 Place the mask over the baby’s mouth and nose, or insert the endotracheal tubeinto de baby’s airway.10 To resuscitate:Place your thumb over the PEEP cap aperture on the T-piece to produce aninspiration.Remove your thumb from the PEEP cap aperture on the T-piece to allow anexpiration.Cleaning.Ensure all oxygen and air supplies are turned off and disconnected from the Neopuffbefore performing cleaning procedures.Clean the Neopuff and accessories either weekly or between babies.Do not use solvents, alcohol-based, or abrasive cleaning solutions.Ensure all parts and accessories are checked before returning the device to service.For ethylene oxide gases: Some carrier gases can cause stress craking and are notSteam autoclavable136 ºC / 220 kPa / 4 minutesSteam autoclavable120 ºC / 96 kPa / 15 minutesNeopuff Test lung Mask Supply tube T-piece✔ ✔ ✔✔ ✔ ✔Ethylene Oxide Gas ✔ ✔ ✔ ✔Chemical liquid immersion ✔ ✔Wipe with soft damp cloth✔MAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050