GUIDO RAYOS X, S.A.25supply tube.Check that the Circuit Pressure gauge reads zero with the Gas inlet disconnected.Occlude the samall aperture in the Positive End Expiratory Pressure (PEEP) capon the T-piece with your thumb.Turn the Inspiratory Pressure (PIP) control clockwise until the manometerneedle stops. This may require few rotations due to the fine thread pitch of thevalve. The manometer now indicates the current Maximum Pressure Reliefsetting.5 To adjust the maximum Pressure Relief:Determine the current Maximum Pressure relief setting as described in step 4 .Occlude the PEEP cap aperture with your thumb.Rotate the Maximum Pressure Relief control to the desired Maximum PressureRelief setting.6 To adjust the Inspiratory Pressure (PIP) valve:Check the test lung is attached to the T-piece and occlude the PEEP cap aperturewith your thumb.Adjust the Inspiratory Pressure contol knob to the desired peal inspiratorypressure as displayed on the manometer.7 To adjust the Positive End Expiratory Pressure (PPEP) valve:Remove your thumb form the PEED cap aperture to allow gas to flow throughMAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050