GUIDO RAYOS X, S.A.171. SENSOR ALARM:In case of air sensor failure or it gets disconnected or out of range; theSENSOR Alarm will activate the corresponding LED and buzzer, thealarm cannot be silenced. The LCD will inform to the attendantdisplaying ,* ALARM *AIR SENSORThe Radiant warmer is now out of service. Call to any authorized Serviceengineer.2. LOW TEMPERATURE ALARM:Is activated when temperature reaches aproximately 1 ºC below the presettemperature (See paragrap II.10 to change preset temperature alarmrange). "LOW" alarm LED will lit, to silence acoustic alarm pressSILENCE key. The LCD will display* ALARM *AT - RH OXThe alarm can be silenced by pressing the SILENCE key. If SILENCEwas pressed and the alarm conditions remain during a period of tenminutes, the alarm will be activated again.3. HIGH TEMPERATUREIs activated when the temperature reaches aproximately 1 ºC higher thanthe preset temperature (See paragrap II.10 to change preset temperaturealarm range). "HIGH" alarm LED will lit and the LCD will display ,* ALARM *AT + RH OXto silence acoustic alarm press SILENCE key. If SILENCE was pressedand the alarm conditions remain during a period of ten minutes, the alarmwill be activated again.Check if heat sources are close to the Radiant warmer location, if positivekeep them away. If the alarm is still effective unplug the unit from thepower source, remove the patient from the radiant warmer and call to theauthorised service engineer.B) SERVO CONTROL MODE1. SENSOR ALARM:MAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!