GUIDO RAYOS X, S.A.161=OXYG, 2=HUMIDNOTE:DUE TO THE WEAR OF THE OXYGEN CELL SENSOR IT ISRECOMMENDED TO PERFORM PERIODICALLY ITS CALIBRATION PROCEDURES.Check the Oxygen Sensor is pluged on the front panel and its housing isnot located inside the Tent Hood, allow to stabilize.Press 1, the LCD will display,AIR=__._ ºCOX: 1=PRG, 2=CALPress 2, the LCD will display ,OX CALIBR (21 %)PRESS ENTERBy pressing ENTER, the Oxygen Sensor is already calibrated with a newreference.New screen displayed ;AIR=__._ ºCOX: 1=PRG, 2=CALII.6. ALARMSThe NESTOMAT 6050 Alarm System informs to the medical attendant of anyincidence may occur. The alarm is visible and audible.NOMENCLATURE OF SYMBOLS IN ALARMMESSAGES+ Excess temperature alarm- Failing temperature alarm= Temperautre O.K.* Desactive sensorA) AIR CONTROL MODEMAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050