GUIDO RAYOS X, S.A.15II.5 OXYGEN MONITORING (OPTIONAL).NESTOMAT 6050 Radiant Warmer monitors the oxygen percentage (optional) whenused with a tent hood (optional).To set the alarm level for above mentioned parameter, proceed as follows (do thecalibration with the sensor out of the tent hood, as it has to be done at a concentration ofoxygen of 21%, which is the oxygen concentration of air) :NOTE: WHEN SELECTING YOU MUST START FROM “SELECT MODE” SCREEN.AIR=__._ ºCSELECT MODEPress the O2/HUM key, the LCD will display ,AIR=__._ ºC1=OXYG, 2=HUMIDIMPORTATNT: OPTION 2=HUMID IS NOT OPERATIVE FOR RADIANT WARMERS.To set the Oxygen range press 1, the LCD will displayAIR=__._ ºCOX: 1=PRG, 2=CALThen, press 1 again ,AT=__._ ºC (__._)PRESET OX %=21.0By means of the keypad set the desired value. Press ENTER key toconfirm.Once again you will find following screen :AIR=__._ ºC1=OXYG, 2=HUMIDFinally press ENTER to confirm values.To calibrate the Oxygen Sensor, press the O2/HUM. key, the LCD willdisplay ;AIR=__._ ºCMAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050