GUIDO RAYOS X, S.A.13AT=__._ ºC (__._)UNIT ACTIVEThe selected heating range can be modified at any time by pressingSTART/STOP and after ENTER key; you will find displayed followingmessage :AIR=__._ ºCSELECT MODEThen set the new temperature by the keypad as previously explained andconfirm by pressing ENTER key. The LCD will display the messagesdescribed above. Then press START key.B) SERVO CONTROL MODE:ATTENTION: BEFORE SELECTING SERVO CONTROL MODEPROCEED TO PREHEAT THE RADIANT WARMER ASEXPLAINED PREVIOUSLY IN THE II.2. TO PREHEATCHAPTER.Before anything display on the screen following message :AIR=__._ ºCSELECT MODEPress SERVO key, the LCD will display ,SKIN=__._ ºCPSET SKIN=__._Set the patient temperature between 20 and 40 ºC by means of thekeypad, confirm by pressing ENTER key.NOTE: WHEN SETTING TEMPERATURE ENTER ALWAYS 3 DIGITS, EVEN INCASE OF NO DECIMALS. AS EXAMPLE PRESS 3, THEN 7, THEN 0 TO SET 37DEGRESS.The LCD will display ,ST=__._ ºC (__._)PRESS STARTMAN-012Sept.00Ed. 1 / Rev. 3NESTOMAT 6050