Brooks® QUANTIM® Low Flow Coriolis Precision Mass Flow ...
Brooks® QUANTIM® Low Flow Coriolis Precision Mass Flow ...
Brooks® QUANTIM® Low Flow Coriolis Precision Mass Flow ...
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Section 4 MaintenanceQmB Series IP40, IP65, IP65XPInstallation and Operation ManualX-CM-QmB-engPart Number: 541B029AHGJune, 20084. Provide an electrical signal to the QUANTIM representing a setpointof 50% of full scale flow and adjust the inlet and outlet pressures tocalibration conditions.5. Verify that the electrical output signal reaches 100% of flow. ForMFC applications vary the setpoint over the 1% to 100% range andverify that the output signal matches the setpoint.a. If possible, connect a flow measurement device (such as arotometer) to the QUANTIM outlet to monitor the actual flowbehavior.6. If the QUANTIM is functioning correctly, the problem may lieelsewhere in the flow system. Re-verify the installation as well as theupstream/downstream fluid system configuration.7. Please contact Brooks Instrument for further assistance introubleshooting. Refer to the back cover of this manual for contactinformation.Figure 4-1 Bench Troubleshooting Circuit4-6