BCT-2000 Software and BCT-128 and BCT-256 User's Guide - Alber
BCT-2000 Software and BCT-128 and BCT-256 User's Guide - Alber
BCT-2000 Software and BCT-128 and BCT-256 User's Guide - Alber
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
15.8. Preamp Offset <strong>and</strong> Gain As-Found1. Complete the Preamp Offset <strong>and</strong> Gain – Before Calibration section of the CalibrationRecord as follows:2. On the main menu, click Diagnostics, then click I/V/IT on the drop-down menu.3. Set the DMM to millivolts.4. Connect the DMM to the CS-<strong>2000</strong> 0 - 200mV jacks.5. Set the Reverse Polarity switch to plus (+) <strong>and</strong> adjust the CS-<strong>2000</strong> for a DMM reading of100mV. Record the voltage under DMM Input.6. Observe the Current Count reading on the I/V/IT dialog box. Record the count underCurrent Count - As-Found.7. Set the Reverse Polarity switch to minus (-) <strong>and</strong> adjust the CS-<strong>2000</strong> for a DMM readingof 100mV. Record the voltage under DMM Input.8. Observe the Current Count reading on the I/V/IT dialog box. Record the count underCurrent Count - As-Found.9. Remove the DMM from the CS-<strong>2000</strong>.64