<strong>Lambda</strong> <strong>20</strong>, <strong>Lambda</strong> <strong>40</strong> UV/Vis SpectrometersOpticsBeam center heightBeam cross-sectionOptical pathlength insample compartmentGrating (Monochromator)Radiation sourcesDetector15 mm above cell holder bottom0.5 mm slit ca. 0.25 mm x 7 mm (width x height)1 mm slit ca. 0.6 mm x 7.5 mm (width x height)2 mm slit ca. 1 mm x 7.5 mm (width x height)4 mm slit ca 2 mm x 7.5 mm (width x height),at focal point of sample <strong>and</strong> reference beam insample compartment121 mmHolographic concave grating with 1053 lines/mmin the centerPrealigned deuterium <strong>and</strong> halogen lampsPhotodiodes(One for the sample beam <strong>and</strong> one for the referencebeam)AbscissaWavelength rangeWavelength accuracyWavelength reproducibilitySpectral b<strong>and</strong>width190 nm to 1100 nm;0 nm for alignment purposes±0.3 nm±0.1 nm0.5 nm, 1 nm, 2 nm, 4 nm(selectable fixed slit widths)88