<strong>Lambda</strong> <strong>20</strong>, <strong>Lambda</strong> <strong>40</strong> UV/Vis SpectrometersOpticsBeam center height15 mm above cell holder bottomBeam cross-section 1 mm slit ca. 0.6 mm x 9 mm (width x height)2 mm slit ca. 1 mm x 9 mm (width x height), atfocal point of sample <strong>and</strong> referencebeam in sample compartmentOptical pathlength insample compartmentGrating (Monochromator)Radiation sourcesDetector121 mmHolographic concave grating with 1053 lines/mmin the centerPrealigned deuterium <strong>and</strong> halogen lampsPhotodiodes(One for the sample beam <strong>and</strong> one for thereference beam)AbscissaWavelength rangeWavelength accuracyWavelength reproducibilitySpectral b<strong>and</strong>widthLamp change190 n to 1100 nm;0 nm for alignment purposes±0.3 nm±0.1 nm1 nm or 2 nm (fixed slit)Automatically at 326 nm(selectable over the whole wavelengthrange, see Super User FactoryConfiguration)Scan speeds 7.5, 15, 30 , 60, 1<strong>20</strong>, 2<strong>40</strong>, 480, 960,19<strong>20</strong> <strong>and</strong> 2880 nm/min84