Lambda 20/Lambda 40 Installation and Maintenance Guide - VTP UP
Lambda 20/Lambda 40 Installation and Maintenance Guide - VTP UP
Lambda 20/Lambda 40 Installation and Maintenance Guide - VTP UP
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Maintenance</strong>2. Using a soft cloth <strong>and</strong> mild laboratory detergent solution, lightly scrub away allforeign material.3. Using a clean cloth dampened with water, rinse the cleaned surfacesthoroughly.4. Dry with lint free cloth or tissue.Sample Compartment WindowWindows are provided with the spectrometer. The window is made of silica <strong>and</strong> maybe used in the entire spectral range of the spectrometer.The window seals the sample compartment <strong>and</strong> thus protects the instrument’s opticsfrom dust <strong>and</strong> fuming or aggressive samples.• Generally, the window should be installed at all times.• The window is an optical component <strong>and</strong> requires the same care <strong>and</strong> h<strong>and</strong>lingas cells.• You can remove the window to clean it. The frame is magnetic <strong>and</strong> can beremoved by h<strong>and</strong>.Windows are most suitably cleaned by wiping them with a soft cloth moistenedwith ethanol.53