<strong>Lambda</strong> <strong>20</strong>, <strong>Lambda</strong> <strong>40</strong> UV/Vis SpectrometersWhen the cell holder has been aligned once, you can take it out <strong>and</strong> reinstall itwithout aligning it again.Minimum Volume ApplicationsTo measure minimum sample volumes, use microcells (offered by PerkinElmer).The minimum sample volume required is a function of the cell internal width orvolume <strong>and</strong> is specified below.Cell TypeCellInternalWidthPathlengthMinimumVolumeRequiredPartNumberHeight ofliquid slightlymore thanheight ofbeam2 mm 1 cm <strong>20</strong>0 µL B0631071(pair)4 mm 1 cm <strong>40</strong>0 µL B0631064(pair)CellVolumePathlengthMinimumVolumeRequiredPartNumberCell windowcompletelyfilled withliquid0.5 µL 0.01 cm 2 µL B06310822.5 µL 0.5 cm 5 µL B06310805 µL 0.1 cm 10 µL B06310835 µL 1.0 cm 10 µL B063108130 µL 1.0 cm 50 µL B063107934