CONTENTS<strong>MFC</strong> <strong>300</strong>1 Product features 31.1 The signal converter with the highest performance ....................................................... 31.2 Options and variants......................................................................................................... 51.3 Signal converter/measuring sensor combination possibilities....................................... 71.4 Measuring principle (single tube) .................................................................................... 72 Technical data 92.1 Technical data................................................................................................................... 92.2 Dimensions and weights ................................................................................................ 202.2.1 Housing ................................................................................................................................. 202.2.2 Mounting plate, field housing ............................................................................................... 212.2.3 Mounting plate, wall-mounted housing ............................................................................... 213 Installation 223.1 Intended use ................................................................................................................... 223.2 Installation specifications .............................................................................................. 223.3 Mounting of the compact version................................................................................... 223.4 Mounting the field housing, remote version .................................................................. 223.4.1 Pipe mounting ....................................................................................................................... 223.4.2 Wall mounting ....................................................................................................................... 233.5 Mounting the wall-mounted housing, remote version .................................................. 243.5.1 Pipe mounting ....................................................................................................................... 243.5.2 Wall mounting ....................................................................................................................... 254 Electrical connections 264.1 Important notes on electrical connection...................................................................... 264.2 Connection diagram ....................................................................................................... 264.3 Connecting power, all housing variants......................................................................... 274.4 Inputs and outputs, overview ......................................................................................... 294.4.1 Combinations of the inputs/outputs (I/Os) ........................................................................... 294.4.2 Description of the CG number .............................................................................................. 304.4.3 Fixed, non-alterable input/output versions.......................................................................... 314.4.4 Alterable input/output versions............................................................................................ 334.5 Laying electrical cables correctly .................................................................................. 345 Notes 352 www.krohne.com 11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>PRODUCT FEATURES 11.1 The signal converter with the highest performanceThe <strong>MFC</strong> <strong>300</strong> is a universal Coriolis mass flow signal converter suitable for a wide range ofapplications and installations. The common platform for the modular hardware allows easyselection for the required output options, and is suitable for mounting in various housingconfigurations.The <strong>MFC</strong> <strong>300</strong> is also suitable for all current and future mass flow sensors. The split architecturesolution for the mass flow family ensures maximum security and redundant back-up ofcalibration parameters, should a failure occur. There is no need for reprogramming after afailed unit is replaced.(signal converter in compact housing)1 Communication with any third party system possible via Foundation Fieldbus, Profibus PA/DP or Modbus2 Intuitive navigation and a wide variety of languages integrated as standard for ease of operation3 Supply voltage: 100...230 VAC (standard) and 24 VDC or 24 VAC/DC (optional)(signal converter in wall-mounted housing)1 Large backlit graphic display with 4 optical buttons to operate the signal converter without having to open the housing2 Any combination of up to 4 inputs and outputs is possible11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com3


1 PRODUCT FEATURES<strong>MFC</strong> <strong>300</strong>Highlights• Modular versions from a basic signal converter to a high-end option with multiple outputoptions• Advanced diagnostic functions• Excellent long term stability• Easy to install and program due to improved user interface• Maximum process safety• Tropicalised and Stainless Steel versions for harsh environmental conditionsIndustries• Water & Wastewater• Chemicals• Power plants• Food & Beverage• Machinery• Oil & Gas• Petrochemicals• Pulp & Paper• PharmaceuticalApplications• Liquids and gases• Slurries and viscous products• Concentration measurement for quality control• Measurement of volume flow• Measurement of density and reference density• Custody transfer loading/unloading• Custody transfer measurements4www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>PRODUCT FEATURES 11.2 Options and variantsModular converter concept(signal converter in compact housing)The <strong>MFC</strong> <strong>300</strong> mass flow signal converter is availablein different variants and offers superiorperformance in any conceivable application.From process control in chemistry, to density andconcentration measurements in the food andbeverage industry, to custody transfer filling andtransport measurements for oil and gas right downto conveyor systems in the pulp and paper industry.Coriolis mass flow measuring systems measure themass and volume flow, the density and thetemperature of liquids and gases. In addition, theconcentration in mixtures and slurries can also bedetermined.Remote design in various versionsThe signal converter in the wall-mounted housing isgenerally used when it is difficult to access themeasuring point or when ambient conditions do notallow the use of the compact version.(signal converter in wall-mounted housing)The signal converter in the 19" rack-mountedhousing is typically built into the central controlroom, away from the harsh ambient conditions thatcan be present at the measuring point.(signal converter in 19" rack-mounted housing)11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com5


1 PRODUCT FEATURES<strong>MFC</strong> <strong>300</strong>Signal converter for any applicationThe basic variant covers many applications,featuring a current output with HART ® ,pulse/frequency output, status output and controlinput.In the modular input/output variant, up to four inputsand outputs can be combined in almost any way. Youcan also select passive or active inputs/outputs.All inputs and outputs are galvanically isolated fromeach other and from the rest of the electronicequipment.(signal converter in field housing)Dual-phase pulse outputs are available for custodytransfer measurements.In addition, the electronics can be equipped withFieldbus functionality (i.e. Foundation Fieldbus,Profibus PA/DP, Modbus, etc.) enablingcommunication to any third party system.For devices without Fieldbus functionality, HART ® isstandard on the first current output.DiagnosticsThe default functionality includes extensive selfdiagnosisof the device, its installation and itsapplication. This is done without any additionalsensors, providing you with valuable informationabout the current status of the device, itsmeasurement and its application. One example isthe 2-phase flow signal which can displayinterferences caused by outgassing.For difficult applications, a toolbox is offered. Thatallows you to record all measuring and diagnosticvalues online over a longer period of time and thenevaluate them using toolbox software.As a result, we offer you peace of mind and reductionof operational and maintenance costs.6www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>PRODUCT FEATURES 11.3 Signal converter/measuring sensor combination possibilitiesMeasuring sensor Measuring sensor + signal converter <strong>MFC</strong> <strong>300</strong>Compact1.4 Measuring principle (single tube)Remote fieldhousingRemote wallmountedhousingRemote rackmountedhousingOPTIMASS 1000 OPTIMASS 1<strong>300</strong> C OPTIMASS 1<strong>300</strong> F OPTIMASS 1<strong>300</strong> W OPTIMASS 1<strong>300</strong> ROPTIMASS 2000 OPTIMASS 2<strong>300</strong> C OPTIMASS 2<strong>300</strong> F OPTIMASS 2<strong>300</strong> W OPTIMASS 2<strong>300</strong> ROPTIMASS <strong>300</strong>0 OPTIMASS 3<strong>300</strong> C OPTIMASS 3<strong>300</strong> F OPTIMASS 3<strong>300</strong> W OPTIMASS 3<strong>300</strong> ROPTIMASS 7000 OPTIMASS 7<strong>300</strong> C OPTIMASS 7<strong>300</strong> F OPTIMASS 7<strong>300</strong> W OPTIMASS 7<strong>300</strong> ROPTIMASS 8000 OPTIMASS 8<strong>300</strong> C OPTIMASS 8<strong>300</strong> F OPTIMASS 8<strong>300</strong> W OPTIMASS 8<strong>300</strong> RStatic meter not energised and with no flow1 Measuring tube2 Drive coil3 Sensor 14 Sensor 2A Coriolis single tube mass flowmeter consists of a single measuring tube 1 a drive coil 2 andtwo sensors (3 and 4) that are positioned either side of the drive coil.11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com7


1 PRODUCT FEATURES<strong>MFC</strong> <strong>300</strong>Energised meter1 Measuring tubes2 Direction of oscilation3 Sine waveWhen the meter is energised, the drive coil vibrates the measuring tube causing it to oscillateand produce a sine wave 3. The sine wave is monitored by the two sensors.Energised meter with process flow1 Process flow2 Sine wave3 Phase shiftWhen a fluid or gas passes through the tube, the coriolis effect causes a phase shift in the sinewave that is detected by the two sensors. This phase shift is directly proportional to the massflow.Density measurement is made by evaluation of the frequency of vibration and temperaturemeasurement is made using a Pt500 sensor.8www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 22.1 Technical data• The following data is provided for general applications. If you require data that is morerelevant to your specific application, please contact us or your local representative.• Additional information (certificates, special tools, software,...) and complete productdocumentation can be downloaded free of charge from the website (Download Center).Measuring systemMeasuring principleApplication rangeCoriolis principleMeasurement of mass flow, density, temperature, volume flow, flow velocity,concentrationDesignModular constructionThe measuring system consists of a measuring sensor and a signal converter.Measuring sensorOPTIMASS 1000DN15…50 / ½…2"OPTIMASS 2000 DN100...250 / 4...10"OPTIMASS <strong>300</strong>0 DN01…04 / 1/25...4/25"OPTIMASS 7000DN06…80 / ¼…3"OPTIMASS 8000DN15…100 / ½…4"All sensors are also available in Ex-versions.Signal converterCompact version (C) OPTIMASS x<strong>300</strong> C (x = 1, 2, 3, 7 or 8)Field housing (F) -<strong>MFC</strong> <strong>300</strong> Fremote versionWall-mounted housing (W) - <strong>MFC</strong> <strong>300</strong> Wremote version19" rack-mounted housing (R) - <strong>MFC</strong> <strong>300</strong> Rremote versionCompact and field housing versions are also available in Ex versions.OptionsOutputs / inputsCountersVerificationConcentration measurementCurrent- (incl. HART ® ), pulse, frequency, and/or status output, limit switch and/orcontrol input (depending on the I/O version)2 (optional 3) internal counters with a max. of 8 counter places (e.g. for countingvolume and/or mass units)Integrated verification, diagnostic functions: measuring device, process, measuredvalue, stabilizationConcentration and concentration flowCommunication interfaces Foundation Fieldbus, Profibus PA and DP, Modbus, HART ®11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com9


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>Display and user interfaceGraphic displayOperating elementsRemote controlDisplay functionsOperating menuLanguage display texts (aslanguage package)Measurement functionsDiagnostic functionsMeasuring accuracyReference conditionsMaximum measuring errorRepeatabilityLC display, backlit white.Size: 128 x 64 pixels, corresponds to 59 x 31 mm = 2.32" x 1.22"Display can be turned in steps of 90°.Ambient temperatures below -25°C / -13°F, may affect the readability of the display.4 optical keys for operator control of the signal converter without opening thehousing.Infrared interface for reading and writing all parameters with IR interface (option)without opening the housing.PACTware ® (incl. Device Type Manager (DTM))HART ® Hand Held Communicator from Emerson ProcessAMS ® from Emerson ProcessPDM ® from SiemensAll DTMs and drivers are available free of charge from the manufacturer's website.Setting the parameters using 2 measured value pages, 1 status page, 1 graphicspage (measured values and graphics are freely adjustable)Standard: English, French, German, Dutch, Portuguese, Swedish, Spanish, ItalianEastern Europe (in preparation): English, Slovenian, Czech, HungarianNorthern Europe (in preparation): English, Danish, PolishChina (in preparation): English, ChineseRussia: English, RussianUnits: Metric, British and US units selectable as desired from lists for volume/massflow and counting, velocity, temperature, pressureMeasured values: Mass flow, total mass, temperature, density, volume flow, totalvolume, velocity, flow direction (not displayed unit – but available via outputs), BRIX,Baume, NaOH, Plato, API, mass concentration, volume concentrationStandards: according to VDI / NAMUR / WIB 2650 (pending) and functions goingbeyond thatStatus messages: Output of status messages optional via display, current and/orstatus output, HART® or bus interfaceSensor diagnostics: Sensor values, drive level, measuring tube frequency, MT(measuring tube) strain, IC (inner cylinder) strain, sensor electronics/boardelectronics temperature, 2-phase flow signalMedium: waterTemperature: 20°C / 68°FPressure: 1 bar / 14.5 psi±0.10% of the measured value ± zero point stability (depending on the measuringsensor)Current output electronics: ±5 µA±0.05% ± zero point stability (depending on the measuring sensor)10www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 2Operating conditionsTemperatureProcess temperatureAmbient temperatureStorage temperaturePressureMediumAmbient pressureChemical propertiesPhysical conditionFlow rateOther conditionsProtection category acc. toIEC 529 / EN 60529Installation conditionsInstallationDimensions and weightsRefer to technical data for the measuring sensor.Depends on the version and combination of outputs.It is a good idea to protect the converter from external heat sources such as directsunlight as higher temperatures reduce the life cycle of all electronic components.-40…+65°C / -40…+149°FStainless Steel housing: -40…+55°C / -40…+131°FAmbient temperatures below -25°C / -13°F, may affect the readability of the display.-50…+70°C / -58…+158°FRefer to technical data for the measuring sensor.AtmosphereLiquids, gases and slurriesRefer to technical data for the measuring sensor.C (compact version) & F (field housing):IP66/67 (acc. to NEMA 4/4X)W (wall-mounted housing):IP 65 (acc. to NEMA 4/4x)R (19" rack-mounted housing):IP 20 (acc. to NEMA 1)For detailed information, refer to chapter "Installation conditions".For detailed information refer to chapter "Dimensions and weights".11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com11


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>MaterialsSignal converter housingMeasuring sensorElectrical connectionGeneralPower supplyPower consumptionSignal cableCable entriesStandardVersion C and F: die-cast Aluminium (polyurethane-coated)Version W: polyamide - polycarbonateVersion R: Aluminium, Stainless Steel and Aluminium sheet, partially polyestercoatedOptionVersions C and F: Stainless Steel 316 L (1.4408)See the technical data for the measuring sensor for housing materials, processconnections, measuring tubes, accessories and gaskets.Electrical connection is carried out in conformity with the VDE 0100 directive"Regulations for electrical power installations with line voltages up to 1000 V" orequivalent national regulations.Standard: 100…230 VAC (-15% / +10%), 50/60 HzOption 1: 24 VDC (-55% / +30%)Option 2: 24 VAC/DC (AC: -15% / +10%, 50/60 Hz; DC: -25% / +30%)AC: 22 VADC: 12 WOnly for remote versions.4 core shielded cable. Detailed specifications available on request.Length: max. <strong>300</strong> m / 1000 ftStandard: M20 x 1.5 (8...12 mm)Option: ½" NPT, PF ½12www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 2Inputs and outputsGeneralDescription of used abbreviationsAll outputs are electrically isolated from each other and from all other circuits.All operating data and output values can be adjusted.U ext = external voltage; R L = load + resistance;U o = terminal voltage; I nom = nominal currentSafety limit values (Ex-i):U i = max. input voltage; I i = max. input current; P i = max. input power rating;C i = max. input capacity; L i = max. input inductivityCurrent outputOutput data Volume flow, mass flow, temperature, density, flow velocity, diagnostic value, 2-phase signalConcentration and concentration flow are also possible with available concentrationmeasurement (optional).Temperature coefficientTypically ±30 ppm/KSettingsWithout HART ®Q = 0%: 0…20 mA; Q = 100%: 10…20 mAError identification: 3…22 mAOperating dataActiveWith HART ®Q = 0%: 4…20 mA; Q = 100%: 10…20 mAError identification: 3…22 mABasic I/OsU int, nom = 24 VDCModular I/OsEx iU int, nom = 20 VDCPassiveI ≤ 22 mAR L ≤ 1kΩU ext ≤ 32 VDCI ≤ 22 mAU 0 ≤ 1.8 VR L ≤ (U ext -U 0 )/I maxI ≤ 22 mAR L ≤ 450 ΩU 0 = 21 VI 0 =90mAP 0 =0.5WC 0 = 90 nF / L 0 = 2 mHC 0 =110nF/ L 0 =0.5mHU ext ≤ 32 VDCI ≤ 22 mAU 0 ≤ 4 VR L ≤ (U ext -U 0 )/I maxU i = 30 VI i = 100 mAP i =1WC i =10nFL i ~0mH11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com13


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>HART ®DescriptionLoadMultidrop operationDevice driverRegistration (HARTCommunication Foundation)Pulse or frequency outputOutput dataFunctionPulse rate/frequencySettingsOperating dataHART ® protocol via active and passive current outputHART ® version: V5Universal HART ® parameter: completely integrated≥ 250 Ω at HART ® test point;Note maximum load for current output!Yes, current output = 4 mAMultidrop address adjustable in operation menu 1…15Available for FC 375, AMS, PDM, FDT/DTMYesPulse output: volume flow, mass flow, mass or volume of dissolved substance withactivated concentration measurementFrequency output: flow velocity, mass flow, temperature, density, diagnostic valueOptional: concentration, flow of the dissolved substanceAdjustable as pulse or frequency output0.01...10000 pulses/s or HzMass or volume per pulse or max. frequency for 100% flowPulse width: setting automatic, symmetric or fixed (0.05...2000 ms)Basic I/OsModular I/OsEx iActive - U nom = 24 VDC -f max in operating menu settof max ≤ 100 Hz:I ≤ 20 mAopen:I ≤ 0.05 mAclosed:U 0, nom = 24 Vat I=20mAf max in operating menu setto100 Hz < f max ≤ 10 kHz:I ≤ 20 mAopen:I ≤ 0.05 mAclosed:U 0, nom = 22.5 Vat I=1mAU 0, nom = 21.5 Vat I=10mAU 0, nom = 19 Vat I=20mA14www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 2Passive U ext ≤ 32 VDC -f max in operating menu set tof max ≤ 100 Hz:I ≤ 100 mAopen:I ≤ 0.05 mA at U ext = 32 VDCclosed:U 0, max = 0.2 V at I ≤ 10 mAU 0, max = 2 V at I ≤ 100 mAf max in operating menu set to100 Hz < f max ≤ 10 kHz:I ≤ 20 mAopen:I ≤ 0.05 mA at U ext =32VDCclosed:U 0, max = 1.5 V at I ≤ 1mAU 0, max = 2.5 V at I ≤ 10 mAU 0, max = 5.0 V at I ≤ 20 mANAMUR - Passive to EN 60947-5-6open:I nom =0.6mAPassive to EN 60947-5-6open:I nom =0.43mALow flow cut-offFunctionclosed:I nom =3.8mAclosed:I nom =4.5mAU i = 30 VI i = 100 mAP i =1WC i =10nFL i ~0mHSwitching point and hysteresis separately adjustable for each output, counter andthe displaySwitching point Set in increments of 0.1.0…20% (current output, frequency output)Hysteresis Set in increments of 0.1.0…5% (current output, frequency output)Time constantFunctionThe time constant corresponds to the elapsed time until 67% of the end value hasbeen reached according to a step function.Settings Set in increments of 0.1.0…100 s11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com15


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>Status output / limit switchFunction and settingsOperating dataAdjustable as automatic measuring range conversion, display of flow direction,overflow, error, switching pointValve control with activated dosing functionStatus and/or control: ON or OFFBasic I/OsModular I/OsActive - U int = 24 VDCI ≤ 20 mAopen:I ≤ 0.05 mAEx i-closed:U 0, nom = 24 Vat I=20mAPassiveU ext ≤ 32 VDCU ext = 32 VDC-I ≤ 100 mAI ≤ 100 mAopen:I ≤ 0.05 mA atU ext =32VDCclosed:U 0, max = 0.2 Vat I ≤ 10 mAU 0, max = 2 Vat I ≤ 100 mAR L, max = 47 kΩR L, min =(U ext -U 0 )/I maxopen:I ≤ 0.05 mAat U ext =32VDCclosed:U 0, max = 0.2 Vat I ≤ 10 mAU 0, max = 2 Vat I ≤ 100 mANAMUR - Passive to EN 60947-5-6open:I nom =0.6mAclosed:I nom =3.8mAPassive to EN 60947-5-6open:I nom =0.43mAclosed:I nom =4.5mAU i = 30 VI i = 100 mAP i =1WC i =10nFL i = 0 mH16www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 2Control inputFunctionHold output values (e.g. when cleaning), set value of outputs to "zero", counter anderror reset, hold counter, range conversion, zero point calibration.Start of dosing when dosing function is activated.Operating dataBasic I/OsModular I/OsEx iActive - U int = 24 VDC-Ext. contact open:U 0, nom = 22 VExt. contact closed:I nom = 4 mAContact closed (on):U 0 ≥ 12 Vwith I nom =1.9mAContact open (off):U 0 ≤ 10 Vwith I nom =1.9mAPassive8V≤ U ext ≤ 32 VDC3V≤ U ext ≤ 32 VDCU ext ≤ 32 VDCI max = 6.5 mAat U ext ≤ 24 VDCI max = 8.2 mAat U ext ≤ 32 VDCContact closed (on):U 0 ≥ 8 Vwith I nom = 2.8 mAContact open (off):U 0 ≤ 2.5 Vwith I nom = 0.4 mAI max = 9.5 mAat U ext ≤ 24 VI max = 9.5 mAat U ext ≤ 32 VContact closed (on):U 0 ≥ 3 Vwith I nom =1.9mAContact open (off):U 0 ≤ 2.5 Vwith I nom =1.9mANAMUR - Active to EN 60947-5-6I ≤ 6 mA at U ext =24VI ≤ 6.6 mA at U ext =32VOn:U 0 ≥ 5.5 V or I ≥ 4mAOff:U 0 ≤ 3.5 V or I ≤ 0.5 mAU i = 30 VI i = 100 mAP i =1WC i = 10 nFL i = 0 mH-Terminals open:U 0, nom = 8.7 VContact closed (on):U 0, nom = 6.3 Vwith I nom > 1.9 mAContact open (off):U 0, nom = 6.3 Vwith I nom < 1.9 mADetection of cable break:U 0 ≥ 8.1 Vwith I ≤ 0.1 mADetection of cable shortcircuit:U 0 ≤ 1.2 Vwith I ≥ 6.7 mA11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com17


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>PROFIBUS DPDescription Galvanically isolated acc. to IEC 61158Profile version: 3.01Automatic data transmission rate recognition (max. 12 MBaud)Bus address adjustable via local display at the measuring deviceFunction blocks8 x analogue input, 3 x totaliserOutput dataMass flow, volume flow, mass counter 1 + 2, volume counter, product temperature,several concentration measurements and diagnostic dataPROFIBUS PADescription Galvanically isolated acc. to IEC 61158Profile version: 3.01Current consumption: 10.5 mAPermissible bus voltage: 9…32 V; in Ex application: 9...24 VBus interface with integrated reverse polarity protectionTypical error current FDE (Fault Disconnection Electronic): 4.3 mABus address adjustable via local display at the measuring deviceFunction blocks8 x analogue input, 3 x totaliserOutput dataMass flow, volume flow, mass counter 1 + 2, volume counter, product temperature,several concentration measurements and diagnostic dataFOUNDATION FieldbusDescription Galvanically isolated acc. to IEC 61158Current consumption: 10.5 mAPermissible bus voltage: 9…32 V; in Ex application: 9...24 VBus interface with integrated reverse polarity protectionLink Master function (LM) supportedTested with Interoperable Test Kit (ITK) version 5.1Function blocks6 x analogue Input, 3 x integrator, 1 x PIDOutput dataMass flow, volume flow, density, temperature of tube, several concentrationmeasurements and diagnostic dataModbusDescriptionAddress range 1…247Modbus RTU, Master / Slave, RS485Supported function codes 01, 03, 04, 05, 08, 16Broadcast Supported with function code 16Supported Baudrate1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud18www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 2Approvals and certificatesCEThe device fulfils the statutory requirements of the EC directives. The manufacturercertifies that these requirements have been met by applying the CE marking.Non-ExStandardHazardous areasOption (only version C)ATEXII 2 G Ex d [ib] IIC T6....T1II 2 G Ex de [ib] IIC T6....T1II 2 D Ex tD A21 IP6x T160°C (dep. on the measuring sensor)without heating jacket or sensor insulationII 2 D Ex tD A21 IP6x T170°C (dep. on the measuring sensor)with heating jacket and sensor insulationII 2(1) G Ex d [ia/ib] IIC T6....T1II 2(1) G Ex de [ia/ib] IIC T6....T1II 2(1) D Ex tD [iaD] A21 IP6x T160°C (dep. on the measuring sensor)without heating jacket or sensor insulationII 2(1) D Ex tD [iaD] A21 IP6x T170°C (dep. on the measuring sensor)with heating jacket and sensor insulationOption (only version F)ATEXII 2 G Ex d [ib] IIC T6II 2 G Ex de [ib] IIC T6II 2(1) G Ex d [ia/ib] IIC T6II 2(1) G Ex de [ia/ib] IIC T6II 2 D Ex tD [ibD] A21 IP6x T80°CII 2(1) G Ex tD [iaD/ibD] A21 IP6x T80°CNepsiEx de ib [ia/ib] IIC T6Ex d ib [ia/ib] IIC T6Optional (only versions C and F)FM / CSAClass I, Div 1 groups B, C, DClass II, Div 1 groups E, F, GClass III, Div 1 hazardous areasClass I, Div 2 groups B, C, DClass II, Div 2 groups F, GClass III, Div 2 hazardous areasIECEx (pending) Ex zone 1 + 2TIIS (in preparation) Zone 1/2Custody transferWithoutStandardOption Liquids other than water 2004/22/EC (MID) acc. to OIML R 117-1Other standards and approvalsShock and vibration resistance IEC 68-2-3Electromagnetic compatibility 2004/108/EC in conjunction with EN 61326-1 (A1, A2)(EMC)European Pressure Equipment PED 97/23 (only for compact versions)DirectiveNAMUR NE 21, NE 43, NE 5311/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com19


2 TECHNICAL DATA<strong>MFC</strong> <strong>300</strong>2.2 Dimensions and weights2.2.1 Housing1 Compact version (C)2 Field housing (F) - remote version3 Wall-mounted housing (W) - remote version4 19" rack-mounted housing (R) - remote versionDimensions and weights in mm and kgVersion Dimensions [mm] Weight [kg]Dimensions and weights in inches and lba b c d e g hC 202 120 155 260 137 - - 4.2F 202 120 155 - - 295.8 277 5.7W 198 138 299 - - - - 2.4R 142(28 TE)129(3 HE)195 - - - - 1.2Version Dimensions [inches] Weight [lb]a b c d e g hC 7.75 4.75 6.10 10.20 5.40 - - 9.30F 7.75 4.75 6.10 - - 11.60 10.90 12.60W 7.80 5.40 11.80 - - - - 5.30R 5.59(28 TE)5.08(3 HE)7.68 - - - - 2.6520www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>TECHNICAL DATA 22.2.2 Mounting plate, field housingDimensions in mm and inches[mm][inches]a 60 2.4b 100 3.9c Ø9 Ø0.42.2.3 Mounting plate, wall-mounted housingDimensions in mm and inches[mm][inches]a Ø9 Ø0.4b 64 2.5c 16 0.6d 6 0.2e 63 2.5f 4 0.2g 64 2.5h 98 3.8511/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com21


3 INSTALLATION<strong>MFC</strong> <strong>300</strong>3.1 Intended useThe mass flowmeters are designed exclusively to directly measure mass flow rates, productdensity and temperature as well to indirectly measure parameters such as the total volume andconcentration of dissolved substances as well as the volume flow rate.For devices used in hazardous areas, additional safety notes apply; please refer to the Exdocumentation.3.2 Installation specificationsThe following precautions must be taken to ensure reliable installation.• Make sure that there is adequate space to the sides.• Protect the signal converter from direct sunlight and install a sun shade if necessary.• Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heatexchanger.• Do not expose the signal converter to intense vibration. The flowmeters are tested for avibration level in accordance with IEC 68-2-3.3.3 Mounting of the compact versionThe signal converter is mounted directly on the measuring sensor. For installation of theflowmeter, please observe the instructions in the supplied product documentation for themeasuring sensor.3.4 Mounting the field housing, remote versionAssembly materials and tools are not part of the delivery. Use the assembly materials and toolsin compliance with the applicable occupational health and safety directives.3.4.1 Pipe mountingFigure 3-1: Pipe mounting of the field housing1 Fix the signal converter to the pipe.2 Fasten the signal converter using standard U-bolts and washers.3 Tighten the nuts.22www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>INSTALLATION 33.4.2 Wall mountingFigure 3-2: Wall mounting of the field housing1 Prepare the holes with the aid of the mounting plate. For further information refer to Mountingplate, field housing on page 21.2 Use the mounting material and tools in compliance with the applicable occupational healthand safety directives.3 Fasten the housing securely to the wall.Mounting multiple devices next to each othera ≥ 600 mm / 23.6"b ≥ 250 mm / 9.8"11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com23


3 INSTALLATION<strong>MFC</strong> <strong>300</strong>3.5 Mounting the wall-mounted housing, remote versionAssembly materials and tools are not part of the delivery. Use the assembly materials and toolsin compliance with the applicable occupational health and safety directives.3.5.1 Pipe mountingFigure 3-3: Pipe mounting of the wall-mounted housing1 Fasten the mounting plate to the pipe with standard U-bolts, washers and fastening nuts.2 Screw the signal converter to the mounting plate with the nuts and washers.24www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>INSTALLATION 33.5.2 Wall mountingFigure 3-4: Wall mounting of the wall-mounted housing1 Prepare the holes with the aid of the mounting plate. For further information refer to Mountingplate, wall-mounted housing on page 21.2 Fasten the mounting plate securely to the wall.3 Screw the signal converter to the mounting plate with the nuts and washers.Mounting multiple devices next to each othera ≥ 240 mm / 9.4"11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com25


4 ELECTRICAL CONNECTIONS<strong>MFC</strong> <strong>300</strong>4.1 Important notes on electrical connectionElectrical connection is carried out in conformity with the VDE 0100 directive "Regulations forelectrical power installations with line voltages up to 1000 V" or equivalent national regulations.• Use suitable cable entries for the various electrical cables.• The sensor and converter are configured together in the factory. For this reason, pleaseconnect the devices in pairs.4.2 Connection diagramThe device must be grounded in accordance with regulations in order to protect personnelagainst electric shocks.Figure 4-1: Connection diagram for remote versions, wall, field and 19" rack-mounted version1 Terminal compartment for signal converter2 Terminal compartment for measuring sensor3 Connect shielding to spring terminal4 Connect shielding to terminal S(with 19" rack-mounted housing, the shielding can be connected to 22z, 22d, 24z or 24d)5 Functional ground26www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>ELECTRICAL CONNECTIONS 44.3 Connecting power, all housing variantsThe device must be grounded in accordance with regulations in order to protect personnelagainst electric shocks.For devices used in hazardous areas, additional safety notes apply; please refer to the Exdocumentation.• The protection category depends on the housing versions (IP65...67 to IEC 529 / EN 60529 orNEMA4/4X/6).• The housings of the devices, which are designed to protect the electronic equipment fromdust and moisture, should be kept well closed at all times. Creepage distances andclearances are dimensioned to VDE 0110 and IEC 664 for pollution severity 2. Supply circuitsare designed for overvoltage category III and the output circuits for overvoltage category II.• Fuse protection (I N ≤ 16 A) for the infeed power circuit, and also a disconnecting device(switch, circuit breaker) to isolate the signal converter must be provided.Power supply connection (excluding 19" rack-mounted housing)1 100...230 VAC (-15% / +10%)2 24 VDC (-55% / +30%)3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%)11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com27


4 ELECTRICAL CONNECTIONS<strong>MFC</strong> <strong>300</strong>100...230 VAC (tolerance range: -15% / +10%)• Note the power supply voltage and frequency (50...60 Hz) on the nameplate.• The protective ground terminal PE of the power supply must be connected to the separate U-clamp terminal in the terminal compartment of the signal converter240 VAC+5% is included in the tolerance range.24 VDC (tolerance range: -55% / +30%)24 VAC/DC (tolerance ranges: AC: -15% / +10%; DC: -25% / +30%)• Note the data on the nameplate!• For measurement process reasons, a functional ground FE must be connected to theseparate U-clamp terminal in the terminal compartment of the signal converter.• When connecting to functional extra-low voltages, provide a facility for protective separation(PELV) (as per VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or relevant national regulations).For 24 VDC, 12 VDC-10% is included in the tolerance range.Power supply connection for 19" rack-mounted housing28www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>ELECTRICAL CONNECTIONS 44.4 Inputs and outputs, overview4.4.1 Combinations of the inputs/outputs (I/Os)This signal converter is available with various in-/output combinations.Basic version• Has 1 current, 1 pulse and 2 status outputs / limit switches.• The pulse output can be set as status output/limit switch and one of the status outputs as acontrol input.Ex i version• Depending on the task, the device can be configured with various output modules.• Current outputs can be active or passive.• Optionally available also with Foundation Fieldbus and Profibus PAModular version• Depending on the task, the device can be configured with various output modules.Bus systems• The device allows intrinsically safe and non intrinsically safe bus interfaces in combinationwith additional modules.• For connection and operation of bus systems, please note the separate documentation.Ex option• For hazardous areas, all of the input/output variants for the housing designs C and F withterminal compartment in the Ex-d (pressure-resistant casing) or Ex-e (increased safety)versions can be delivered.• Please refer to the separate instructions for connection and operation of the Ex-devices.11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com29


4 ELECTRICAL CONNECTIONS<strong>MFC</strong> <strong>300</strong>4.4.2 Description of the CG numberFigure 4-2: Marking (CG number) of the electronics module and input/output variants1 ID number: 22 ID number: 0 = standard; 9 = special3 Power supply option4 Display (language versions)5 Input/output version (I/O)6 1st optional module for connection terminal A7 2nd optional module for connection terminal BThe last 3 digits of the CG number (5, 6 und 7) indicate the assignment of the terminalconnections. Please see the following examples.Examples for CG numberCG 320 11 100CG 320 11 7FKCG 320 81 4EB100...230 VAC & standard display; basic I/O: I a or I p & S p /C p & S p & P p /S p100...230 VAC & standard display; modular I/O: I a & P N /S N and optional module P N /S N & C N24 VDC & standard display; modular I/O: I a & P a /S a and optional module P p /S p & I pDescription of abbreviations and CG identifier for possible optional moduleson terminals A and BAbbreviation Identifier for CG No. DescriptionI a A Active current outputI p B Passive current outputP a / S a C Active pulse, frequency, status output or limit switch (changeable)P p / S p E Passive pulse, frequency, status output or limit switch (changeable)P N / S N F Passive pulse, frequency, status output or limit switch according toNAMUR (changeable)C a G Active control inputC p K Passive control inputC N H Active control input to NAMURSignal converter monitors cable breaks and short circuits acc. toEN 60947-5-6. Errors indicated on LC display. Error messages possiblevia status output.IIn a P Active current inputIIn p R Passive current input- 8 No additional module installed- 0 No further module possible30www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>ELECTRICAL CONNECTIONS 44.4.3 Fixed, non-alterable input/output versionsThis signal converter is available with various in-/output combinations.• The grey boxes in the tables denote unassigned or unused connection terminals.• In the table, only the final digits of the CG-No. are depicted.• Connection terminal A+ is only operable in the basic input/output version.CG-No.Connection terminalsA+ A A- B B- C C- D D-Basic in-/output (I/O) (Standard)1 0 0 I p + HART ® passive 1 S p / C p passive 2 S p passive P p / S p passive 2I a + HART ® active 1Ex-i inputs/outputs (optional)2 0 0 I a + HART ® active P N / S N NAMUR 23 0 0 I p + HART ® passive P N / S N NAMUR 22 1 0 I a active P N / S N NAMURC p passive 23 1 0 I a active P N / S N NAMURC p passive 22 2 0 I p passive P N / S N NAMURC p passive 23 2 0 I p passive P N / S N NAMURC p passive 2I a + HART ® active P N / S N NAMUR 2I p + HART ® passive P N / S N NAMUR 2I a + HART ® active P N / S N NAMUR 2I p + HART ® passive P N / S N NAMUR 2PROFIBUS PA (Ex-i) (Option)D 0 0 PA+ PA- PA+ PA-D 1 0 I a active P N / S N NAMURC p passive 2D 2 0 I p passive P N / S N NAMURC p passive 2FISCO DeviceFISCO DevicePA+ PA- PA+ PA-FISCO DeviceFISCO DevicePA+ PA- PA+ PA-FISCO DeviceFISCO Device11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com31


4 ELECTRICAL CONNECTIONS<strong>MFC</strong> <strong>300</strong>FOUNDATION Fieldbus (Ex-i) (Option)E 0 0 V/D+ V/D- V/D+ V/D-E 1 0 I a active P N / S N NAMURC p passive 2E 2 0 I p passive P N / S N NAMURC p passive 21 function changed by reconnecting2 changeableFISCO DeviceFISCO DeviceV/D+ V/D- V/D+ V/D-FISCO DeviceFISCO DeviceV/D+ V/D- V/D+ V/D-FISCO DeviceFISCO Device32www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>ELECTRICAL CONNECTIONS 44.4.4 Alterable input/output versionsThis signal converter is available with various in-/output combinations.• The grey boxes in the tables denote unassigned or unused connection terminals.• In the table, only the final digits of the CG-No. are depicted.• Term. = (connection) terminalCG-No.Connection terminalsA+ A A- B B- C C- D D-Modular inputs/outputs (option)4 _ _ max. 2 optional modules for term. A + B I a + HART ® active P a / S a active 18 _ _ max. 2 optional modules for term. A + B I p + HART ® passive P a / S a active 16 _ _ max. 2 optional modules for term. A + B I a + HART ® active P p / S p passive 1B _ _ max. 2 optional modules for term. A + B I p + HART ® passive P p / S p passive 17 _ _ max. 2 optional modules for term. A + B I a + HART ® active P N / S N NAMUR 1C _ _ max. 2 optional modules for term. A + B I p + HART ® passive P N / S N NAMUR 1PROFIBUS PA (Option)D _ _ max. 2 optional modules for term. A + B PA+ (2) PA- (2) PA+ (1) PA- (1)FOUNDATION Fieldbus (Option)E max. 2 optional modules for term. A + B V/D+ (2) V/D- (2) V/D+ (1) V/D- (1)PROFIBUS DP (Option)F _ 01 optional module forterm. ATerminationPRxD/TxD-P(2)RxD/TxD-N(2)TerminationNRxD/TxD-P(1)RxD/TxD-N(1)Modbus (Option)G _ _ 2 max. 2 optional modules for term. A + B Common Sign. B(D1)H _ _ 3 max. 2 optional modules for term. A + B Common Sign. B(D1)1 changeable2 not activated bus terminator3 activated bus terminatorSign. A(D0)Sign. A(D0)11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com33


4 ELECTRICAL CONNECTIONS<strong>MFC</strong> <strong>300</strong>4.5 Laying electrical cables correctlyFigure 4-3: Protect housing from dust and water1 Lay the cable in a loop just before the housing.2 Tighten the screw connection of the cable entry securely.3 Never mount the housing with the cable entries facing upwards.4 Seal cable entries that are not needed with a plug.34www.krohne.com11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en


<strong>MFC</strong> <strong>300</strong>NOTES 511/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 enwww.krohne.com35


KKKKROHNE product overview© KROHNE 11/2009 - 4000498402 - TD <strong>MFC</strong> <strong>300</strong> R03 en - Subject to change without notice.• Electromagnetic flowmeters• Variable area flowmeters• Ultrasonic flowmeters• Mass flowmeters• Vortex flowmeters• Flow controllers• Level meters• Temperature meters• Pressure meters• Analysis products• Measuring systems for the oil and gas industry• Measuring systems for sea-going tankersHead Office KROHNE Messtechnik GmbHLudwig-Krohne-Str. 5D-47058 Duisburg (Germany)Tel.:+49 (0)203 301 0Fax:+49 (0)203 301 10389info@krohne.deThe current list of all KROHNE contacts and addresses can be found at:www.krohne.com