12.07.2015 Views

NCRP #147 Examples - Radiation Shielding for Medical Instalations

NCRP #147 Examples - Radiation Shielding for Medical Instalations

NCRP #147 Examples - Radiation Shielding for Medical Instalations

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Example 5.1: Cath Lab Wall• N=25 patients/wk• d=4 m• Uncontrolled, fully-occupiedarea– P = .02 mGy/wk, T = 1• Let’s use Model 2 with theCardiac Angio workloaddistributionConservatively assume30.5 cm diameterimage intensifier, 180°– K 1 sec = Kerma per patient at 1 m scatter, all primary= 38 3.8 mGy/patient tfrom Table 4.7 47radiation stopped byimage receptor3


Example 5.1: Cath Lab Wall• Total weekly unshielded kerma from thissecondary radiation is then• The required barrier transmission is• On Figure C.2. look up required lead thicknessthat gives this transmission4


Figure C.2Example 5.1Cath Lab WallB=0.0034x=1.3 mm Pb5


Example 5.1: XRAYBARR Output6


Example 5.1: XRAYBARR OutputSimply illustrates self-consistency,since the data in Table 4.7 are basedon XRAYBARR results7


Example 5.2: Dedicated Chest Unit• Need to shield an uncontrolled,fully-occupied area: P = .02mGy/wk, T = 1, in the primarybeam (U=1)• N=300 patients/wk• d=3 m• Let’s use Model 2 with the ChestRoom workload distribution– K 1 pri = Kerma per patient at 1 m =1.2 mGy/patient from Table 4.58


Example 5.2: Chest Unit itP Primary Barrier• Total weekly unshielded kerma from thisprimary radiation is then• The required barrier transmission is• Look up required lead thickness that gives thistransmission on Figure B.2.9


Fig B.2 Example 5.2Chest UnitPrimary WallB=0.0005Total primarybarrier thicknessx=2.2 22mm Pb10


Example 5.2: Chest Unit Primary Barrier• Consider the shielding af<strong>for</strong>ded in the chestunit’s image receptor and its structure– x pre = 0.85 mm lead from Table 4.6• There<strong>for</strong>e, the thickness of the lead in thewallneedstobe22mm2.2 – 085mm=140.85 1.4mm lead– Note: this x pre value is a single conservativevalue from the list in Dixon & Simpkin (1998)and not exact <strong>for</strong> the Chest Room workload11


Example 5.2: XRAYBARR Output“More correct”12


Example 5.2: Dedicated Chest Unit• Now shield the secondary wall(U=0), P/T=0.02 mGy/wk• N=300 patients/wk• d=2.1 m• Let’s use Model 2 with theChest Room workloaddistribution– K 1 sec = Secondary kerma perpatient at 1 m = 2.7 ×10–3 fromTable 4.7xAssume F=1,535cm 2 (at 1.83 mSID),90° scatter13


Example 5.2: Chest Unit itS Secondary Barrier• Total weekly unshielded kerma from thissecondary radiation is then• The required barrier transmission is• Look up required lead thickness that gives thistransmission on Figure C.2.14


Figure C.2B=0.11Example 5.2 ChestRoom Sec Wallx=0.42 mm Pb15


Example 5.3 The Radiographic Room16


Example 5.3 The Radiographic Room• Let N = 125 patients/week• Assume the workload distribution and usefactors follow that from the RadiographicRoom from the AAPM-TG9 survey(Simpkin, 1996)17


Example 5.3.1.1 Floor Under Table• Initially, only consider the primaryradiation from the overhead x-ray tubeposition• Assume the tube usage follows the RadRoom/floor other barriers workloaddistribution• The survey found U = 0.89, but let U =1 to be conservative• From Table 4.5, kerma aper patient t= 5.2mGy patient –1 at 1 m <strong>for</strong> thisdistributionx18


Example 5.3.1.1 Floor Under Table• Then unshielded kerma is• To shield to P/T = 0.0202 mGy/wk, needtransmission• Fig B.3 shows the primary transmissionthrough concrete, requiring 107 mmconcretex19


Figure B.3Example 5.3.1.1Rad Room FloorUnder TableB=5.1×10 -4x =107 mmconcrete20


Example 5.3.1.1 Floor Under Table• The image receptor in the radiographic tableand its support provide attenuation,– x pre = 72 mm concrete (Table 4.6)• Then the required floor thickness = 107 –72 = 35 mm concrete.21


Example 5.3.1.1 Floor Under Table• Now lets try an NT/Pd 2 calculation<strong>for</strong> the floor• This will include all the othersources of radiation i in the room(e.g. scatter from exposuresagainst the chest board, cross tablelateral exposures, etc.)x22


Figure 4.6bExample53115.3.1.1Rad Room FloorUnder Tablex =37 mmconcreteNT/Pd 2 = 372 mGy –1 m –223


Example 5.3.1.1 Floor Under TableNTPD2 output24


Example 5.3.1.1 Floor Under Table• So a primary-only calculation l with U=1requires 35 mm concrete, while an NT/Pd 2calculation l (which h includes all othersources) requires 37 mm concrete.• (As expected, the vast majority of exposurebelow the table is due to the primary overtablebeam, and shielding against just theprimary beam is sufficient.)25


Example 5.3.1.2 Floor Not Under Table• Now let’s look at areas beneathroom, away from table. This willreceive secondary radiation only.• Let d = 3 m (vertical distancefrom patient on table)• Consider secondary radiationfrom the Rad Room (all barriers)workload distribution• 90° scatter• From Table 4.7, secondary kermaper patient = 3.4 ×10 -2 mGypatient–1 at 1 m <strong>for</strong> thisdistributionx26


Example 5.3.1.2 Floor Not Under Table• Then unshielded kerma is• To shield to P/T = 002 0.02 mGy/wk, need transmissioni• Fig C.3 shows the secondary transmission throughconcrete, requiring 33 mm concrete27


Figure C.3 Example 5.3.1.2B=4.3×10 -2 Rad Room FloorNot Under Tablex =33 mmconcrete28


Example 5.3.5 Cross-Table Lateral Wall2 m29


Example535Cross-Table 5.3.5 Lateral Wall• This wall needs to shield against– Primary radiation from cross-table exposures (but withU only = 0.09) following the Rad Room/floor otherbarriers workload distribution. We wish to account <strong>for</strong>image receptor attenuation.– Secondary radiation <strong>for</strong> exposures directed down attable (& chest exposures) following the RadRoom/floor other barriers workload distribution• We can only guess at a solution using Model 2(the kerma per patient scheme) since there is noway to combine the two shielded doses with onetransmission curve30


Example535Cross-Table 5.3.5 Lateral Wall• But Model 3 (NT/Pd 2 scheme) should fit thiswell, since the NT/Pd 2 curves account <strong>for</strong> allsources in the room. Using d = 2.8m fromcross-table primary x-ray tube location,1NT 125 patients wk− × 1==7972 − 12Pd 0.02 mGy wk × (2.8m)mGy• Accounting <strong>for</strong> attenuation in the grid andimage receptor, from Fig 4.5b, 0.83 mm leadis required in the wall.−1m231


Figure 4.5bExample 5.3.5Rad RoomCross-TableLateral Wallx =0.83 mmleadNT/Pd 2 = 797 mGy –1 m –232


Example535Cross-Table 5.3.5 Lateral WallSetup XRAYBARR with 2x-ray sources33


Example535Cross-Table 5.3.5 Lateral WallNTPD2 output34


XRAYBARR• Per<strong>for</strong>ms first-principle extension of <strong>NCRP</strong>-49• Microsoft application <strong>for</strong> Windows written inVisual Basic• Freely-distributed shareware atwww.geocities.com/djsimpkin• Comes as a ~3MB zip file. Save zip on localdrive in a new folder. Unzip constituent ~2dozen files to this new folder. Double-clickSETUP.EXE to install program onto yoursystem.35


NTPd2• Per<strong>for</strong>ms calculations and lookup followingthe NT/Pd 2 model.• Microsoft application <strong>for</strong> Windows writtenin Visual Basic• Freely-distributed shareware• Just 4 files (executable, 2 help files, & iconfile). Keep them in a separate folder and noinstallation needed.36

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!