<strong>OPTIFLEX</strong> <strong>1300</strong> CINSTALLATION 2Figure 2-17: Installation recommendations for solids1 We recommend installation without a nozzle. If not, h ≤50 mm / 2¨.2 Radius of the tank, r3 The end of the probe must be more than 300 mm / 12¨ above the tank bottom.4 Empty space (radius, R min ) around the probe.5 The electromagnetic (EM) field generated by the device. It is also the measurement zone of the probe. Make sure thatthe EM field is clear of objects and product flow.6 Ground the tank, the product and the probe (if attached).7 If possible, put the process fitting ≥300 mm / 12¨ from the tank wallClearance between the probe and other objects in the tankProbe typeEmpty space (radius, R min ) around the probe[mm][inches]Single cable Ø4 mm / 0.15¨ 4 300 12Single cable Ø8 mm / 0.3¨ 4 300 12INFORMATION!If the probe is longer than 10 m / 33 ft, we recommend that you do not attach the end of theprobe.09/2010 - 4000049805 - QS <strong>OPTIFLEX</strong> <strong>1300</strong> R08 enwww.krohne.com17