S E C T I O N 4TROUBLESHOOTING AND PREVENTIVE MAINTENANCE4.2 Cleaning the InstrumentUse the following options to clean Nanoductstraps, holders and attachment strap retainersor any other parts that come into contactwith a patient:Option A:1 Soak the straps for 30 minutes in a freshlyprepared 10% dilution of household bleach.2 Rinse soaked straps thoroughly in tap water.3 Allow to air dry.Option B.1 Autoclave the straps for 30 minutes at 121ºC.CAUTION:Do not autoclave the Nanoduct holders (red or black) orthe attachment strap retainer post. Autoclaving willdestroy these parts.35