S E C T I O N 3INTERPRETING THE SWEAT TEST3.4 Initial Sweating RateOne advantage of the unique continuous-flowsensing device of the Nanoduct System is that itallows computation of the initial sweating rate. Thevolume of the sensor channel from the first electrodeto the second electrode contact is preciselyknown, as is the collecting surface area. After thesensor is attached to the arm, the time taken forthis volume to be filled with sweat is measured bythe Induction/Analysis module.Applying an algorithm with the fill time as the solevariable allows a display of the initial sweating ratein the conventionally accepted units of grams persquare meter of skin surface per minute. Thisdatum is available when the first reading is displayedon the continuous record of conductivity.29