S E C T I O N 2SWEAT INDUCTION AND ANALYSIS2.1 Performing the Sweat TestNANODUCT(Date) (Time)Check ControlsIontophoresisSweat AnalysisRecall ReadingIontophoresis8 Activate Iontophoresisa. After completing Check Controls the arrowon the display should indicate Iontophoresison the menu.<strong>Inc</strong>reasingCurrentIontophoresisFullCurrentENTER to ContinueIontophoresisCompleteb. Press ENTER to start iontophoresis. The displaywill show: INCREASING CURRENT. Oncecurrent has reached the full 0.5mA, the displaywill show FULL CURRENT and that level ismaintained for 2 minutes. Once full currentis achieved the Time Bar shows the remainingtime. Current will then decrease to zero(indicated by a short beep). The display willshow: IONTOPHORESIS COMPLETE.c. Press ENTER to return to the main menu.c9 Remove the Red ElectrodeWhile applying pressure to the red electrodewith the index finger, rotate the upper ringapproximately 90° clockwise until the cutoutsalign with the electrode flanges. Thenremove the electrode.CAUTION:Do not disturb or remove the red holder. It mustremain in place during analysis. In addition, leavethe negative (black) electrode, gel, holder andtape-downs in position to provide a ground contactfor initial sweat rate determination.18