S E C T I O N 2SWEAT INDUCTION AND ANALYSIS2.1 Performing the Sweat Testd. Grip the holder and lift it briefly above theskin to equalize strap tension on each sideof the holder, then replace it on the skin surface.Adjust strap tension as needed toensure correct contact.NOTE Attach the strap firmly, but avoid excessivetightness. Correctly applied, the holdershould grip the skin firmly enough to resistmoderately forceful attempts to change itsposition. The surrounding skin areas shouldmove with the holder when it is moved.e. Draw the skin back around the holder toremove any underlying wrinkles.f. Position the negative (black) holder (on thesame limb) in the same manner.6 Insert Pilogels Into the Iontophoretic ElectrodesInsert a gel into each electrode assembly.Press lightly and rotate to ensure an air-freeand moist interface between electrode plateand gel.7 Insert Each Electrode Into Its HolderFit the red (positive) anode into the holderwith the red locking ring and fit theblack (negative) cathode into the holderwith the black locking ring as follows:16