12.07.2015 Views

HCS 2.4/3.0. The countersinkable compression screw.

HCS 2.4/3.0. The countersinkable compression screw.

HCS 2.4/3.0. The countersinkable compression screw.

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Surgical Technique for <strong>HCS</strong> <strong>2.4</strong>/3.01Insert the guide wireInstruments292.622 Guide Wire 1.1 mm with threaded tiporwith trocar, length 150 mm, Stainless Steel292.623 Guide Wire 1.1 mm with trocar tip,length 150 mm, Stainless Steel312.151 Double Drill Guide 2.0/1.1While monitoring with the image intensifier, advance theguide wire through the drill guide into the bone until thethread tip is anchored in the far cortex.Note: Do not forcefully insert the guide wire. This may causeit to bend.2Determine <strong>screw</strong> and thread lengthInstruments03.226.002 Direct Measuring Device for <strong>HCS</strong> –Headless Compression Screw 3.0 mm292.622 Guide Wire 1.1 mm with threaded tiporwith trocar, length 150 mm, Stainless Steel292.623 Guide Wire 1.1 mm with trocar tip,length 150 mm, Stainless SteelSlide the narrow end of the measuring device over the guidewire to the bone.<strong>The</strong> measurement on the measuring device shows the depthof the guide wire in the bone in millimeters and directly theappropriate <strong>screw</strong> length.Note: Only use the guide wire in its original length to ensurecorrect measurement.6 Synthes <strong>HCS</strong> <strong>2.4</strong>/3.0 Instructions for Use

Hooray! Your file is uploaded and ready to be published.