Adjustments & Performance Tests2.5.7 Cal Factor Write Example NI 488.2 Syntaxwrt$ = “TEST EEPROM A READ”ibwrt pwrmtr%, wrt$spfrqcf = “0.00”’50 MHz cal factorstanFreq =“2.00,3.00,4.00,5.00,6.00,7.00,8.00,9.00,10.00,11.00,12.00,13.00,14.00,15.00,16.00,17.00,18.00”‘standard frequency cal factorswrt$ = “test eeprom a calfsp” + spfrqcfibwrt pwrmtr%, wrt$‘load special frequency cal factor into edit bufferwrt$ = “test eeprom a calfst” + stanfreqibwrt pwrmtr%, wrt$‘load standard frequency cal factors into edit bufferwrt$ = “test eeprom a write 0"ibwrt pwrmtr%, wrt$‘write cal factors from edit buffer to sensor EEPROM2.5.8 Linearity Correction FactorsTEST EEPROM A|B CORR dX6Correction factor range: -99 to 99TEST EEPROM A|B CORR?2.5.9 Start/Stop Frequency, <strong>Power</strong> (dBm), Video R+ and R-TEST EEPROM A|B SPEC dX5TEST EEPROM A|B SPEC?2.5.10 Sensor Temperature Compensation (807XXA Only)TEST EEPROM A|B TC CLR|LO|MID|HITEST EEPROM A|B TC CUR?|LO?|MID?|HI?Returns 9 values. Use the “CUR?” to query last temperature data after a calibration.Publication 32089, Rev. A4, February 2002 2-17