<strong>Series</strong> <strong>8650A</strong> <strong>Universal</strong> <strong>Power</strong> <strong>Meters</strong>2.5.3 Cal Factor Table & Frequency QueryEEPROM A|B CALF?Response: whole cal factor table sorted by frequencyEEPROM A|B FREQ?Response: whole frequency table sorted by frequency2.5.4 Standard Frequency Correction Factor Table & StandardFrequencies TableTEST EEPROM A|B CALFST d*TEST EEPROM A|B CALFST?TEST EEPROM A|B FREQST?2.5.5 Standard Start Frequency, Standard FrequencyIncrement, Number of Standard Frequencies, Number ofSpecial FrequenciesTEST EEPROM A|B CALFR d d n nQuery: TEST EEPROM A|B CALFR?Syntax ExamplesNI 488.2wrt$ = “TEST EEPROM A CALFR 2E9,1E9,17,1”ibwrt pwrmtr%, wrt$mNI VISAwrt$= “TEST EEPROM A CALFR 2E9,1E9,17,1”stat = viWrite(sesn, wrt$, Len(wrt$), retCount)Description: This command sets the lowest frequency in the normal frequency correction table, thefrequency increment, the number of normal frequency correction points and the number ofspecial frequency correction points.2.5.6 Special Frequency Correction Factor Table & SpecialFrequencies TableTEST EEPROM A|B CALFSP d*TEST EEPROM A|B CALFSP?TEST EEPROM A|B FREQSP d*TEST EEPROM A|B FREQSP?2-16 Publication 32089, Rev. A4, February 2002