Adjustments & Performance Tests2.4 Calibrator Service GPIB 488.2 Commands2.4.1 Set or Query Cal Factor of the Calibrator for 50 MHz and1 GHz (Option 12 Only)TEST CALIB KB 50M|1G dTEST CALIB KB 50M?|1G?2.4.2 Set or Query Calibrator Output <strong>Power</strong> LevelTEST CALIB POWER dwhere d is the dBm power levelTEST CALIB POWER?2.5 Sensor EEPROM GPIB 488.2 Commands2.5.1 General Sensor DataTEST EEPROM A|B TYPE Nx9TEST EEPROM A|B TYPE?Example: TEST EEPROM A TYPE 80301, 1800101, 0, 59, 18, 15, 4, 01, 0Set sensor A model, serial number, cal location (see below), minute, hour, day, month, year, passwordflag.Model range: 0 to 999999Serial number range: 0 to 99999999Cal location range: 0 to 3 (0 = factory; 1 = field rep; 2 = service center; 3 = customer site)Password flag: 0 for disable, 1 for enable2.5.2 Load Sensor Data to Edit Buffer and Write from Buffer toSensor EEPROMTEST EEPROM A|B READTEST EEPROM A|B WRITE nwhere n = password number in range of 0 to 999999Publication 32089, Rev. A4, February 2002 2-15