Adjustments & Performance TestsRefer to the Calibrator Troubleshooting section of the manual if the Calibrator does not meet theReturn Loss specification.Figure 2-4: Return Loss Setup2.3.2.5 Calibrator Frequency CheckConnect the output of the <strong>8650A</strong> calibrator to the frequency counter. Set the output of the calibrator to50 MHz, +5 dBm and turn on the calibrator. Record the reading of the frequency counter in thePerformance Verification Test Data Sheet. If Option 12 is installed, repeat this test for 1000 MHz.Manual 32089, Rev. A, December 2000 2-9a