<strong>Series</strong> <strong>8650A</strong> <strong>Universal</strong> <strong>Power</strong> <strong>Meters</strong>10. Calculate the Calibrator Output level (P CAL ) using the following formula:where:2 21 12VPCOMP (V - V0) + V 0 - VCAL(Watts)=4R(Calibration Factor)P CAL = calibrator output power reference levelV COMP = previously recorded value in Step 8V 1 = previously recorded value in Step 7V 0 = previously recorded value in Step 6R= 200Ω (assuming HP478A-H75 mount)Calibration factor = value for the thermistor mount at 50 MHz (traceable to the NIST)11. Verify that the P CAL is within the following limits:1 mW ±0.019 mW (0.981 to 1.019 mW)For record purposes, the measured value of P CAL can be entered on the Test Data RecordingSheet located on page 2-10.12. For Option 12, set the calibrator frequency to 1.000 GHz as in step 8 and repeat steps 7 & 9through 12.2-6a Manual 32089, Rev. A, December 2000