Adjustments & Performance TestsEnter the power level reading into the following formula:CFnew = (-10Log10(Pcal*1000)) * 1000Record the new Cal Factor value (CFnew) into the New Cal Factor column of the Calibrator Cal Factortable. Enter the new Calibrator Cal Factor by typing “T” and Enter. Repeat the power measurement toconfirm the new Cal Factor for 50 MHz. If Option 12 is installed, repeat section 2.2.3 using the 1000MHz settings and readings. To enter the new cal factor, type “U” and Enter.Figure 2-1: Calibrator Output Test Setup2.2.4 Display Brightness AdjustmentThe <strong>Series</strong> <strong>8650A</strong> <strong>Universal</strong> <strong>Power</strong> <strong>Meters</strong> LCD display can be adjusted for brightness. The displaybrightness is adjustable in 5% or 10% increments.Procedure1. From the main display screen, press the "Display" softkey.2. Press the "More Options" softkey.3. Use the UP and DOWN arrow buttons to increase or decrease the brightness setting.4. Press the "OK" softkey to return the main menu.Manual 32089, Rev. A, December 2000 2-3b