Voltage/Current Calibrator DIGISTANT® Model 4411 - Burster

Voltage/Current Calibrator DIGISTANT® Model 4411 - Burster Voltage/Current Calibrator DIGISTANT® Model 4411 - Burster

12.07.2015 Views

OPERATION MANUALVoltage/CurrentCalibratorDIGISTANT ® Model 4411© 2012 bursterpräzisionsmesstechnik gmbh & co kgAll rights reservedValid from 24.09.2012Manufacturer:burster präzisionsmesstechnik gmbh & co kgTalstraße 1 - 5 P.O.Box 143276593 Gernsbach 76587 GernsbachGermanyGermanyTel.: (+49) 07224 / 6450Fax: (+49) 07224 / 64588E-Mail: info@burster.comwww.burster.com1003-BA4411EN-5070-091521

OPERATION MANUAL<strong>Voltage</strong>/<strong>Current</strong><strong>Calibrator</strong>DIGISTANT ® <strong>Model</strong> <strong>4411</strong>© 2012 bursterpräzisionsmesstechnik gmbh & co kgAll rights reservedValid from 24.09.2012Manufacturer:burster präzisionsmesstechnik gmbh & co kgTalstraße 1 - 5 P.O.Box 143276593 Gernsbach 76587 GernsbachGermanyGermanyTel.: (+49) 07224 / 6450Fax: (+49) 07224 / 64588E-Mail: info@burster.comwww.burster.com1003-BA<strong>4411</strong>EN-5070-091521


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Exclusion of warranty liability for operating manualsAll information in the present documentation was prepared and compiled with great care andreproduced subject to effective control measures. No warranty is provided for freedom fromerrors. We reserve the right to make technical changes. The present information as well asthe corresponding technical data can change without notice. Reproduction of any part of thisdocumentation or its processing or revision using electronic systems is prohibited without themanufacturer's prior written approval.Components, devices and measured value sensors made by burster praezisionsmesstechnik(hereinafter referred to as "product") are the results of targeted development and meticulousresearch. As of the date of delivery, burster provides a warranty for the proper condition andfunctioning of these products covering material and production defects for the period specifiedin the warranty document accompanying the product. However, burster excludes guarantee orwarranty obligations as well as any liability beyond that for consequential damages caused byimproper use of the product, in particular the implied warranty of success in the market as wellas the suitability of the product for a particular purpose. Furthermore, burster assumes no liabilityfor direct, indirect or incidental damages as well as consequential or other damages arising fromthe provision and use of the present documentation.Page 2


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®EG-KonformitätserklärungEC- Declaration of Conformity according to EN ISO/IEC 17050-1:2004Name des Herstellers:Manufacturer’s Name:burster präzisionsmesstechnik gmbh & co kgAdresse des Herstellers: Talstr. 1-5Manufacturer’s Address:76593 Gernsbach, Germanyerklärt unter alleiniger Verantwortung, dass das gelieferte Produktdeclares under sole responsibility that the product as originally deliveredProduktname:Product Name:Pocket Spannungs-/Stromkalibriergerät DIGISTANT®Handy <strong>Calibrator</strong> DIGISTANT®<strong>Model</strong>lnummer(n) (Typ): <strong>4411</strong><strong>Model</strong>s Number / Type:Produktoptionen:OptionsDiese Erklärung beinhaltet obengenannte Produkte mit allen OptionenThis declaration covers all options of the above product(s)mit den folgenden europäischen Richtlinien übereinstimmt und entsprechend das CE-Zeichen trägt:complies with the requirements of the following applicable European Directives, and carries the CE marking accordingly:2006/95/ECElektrische Betriebsmittel zur Verwendung innerhalb bestimmter SpannungsgrenzenLow <strong>Voltage</strong>Electrical Equipment designed for use within certain voltage limits2004/108/ECElektromagnetische VerträglichkeitEMCElectromagnetic CompatibilityObengenannte Produkte entsprechen folgenden harmonisierten Normen:Above named products conform with the following product standards:Sicherheit: IEC 61010-1:2001 / EN 61010-1:2001 Messkategorie 1 Schutzklasse III; *Safety requirements: CAT 1 Safety class 3* external AC adapter (220-240 V) 8 V DC / 150 mA <strong>Model</strong> <strong>4411</strong>-Z001 Schutzklasse 2Safety class 2EMV Störaussendung: IEC/CISPR 11:2003 + A1:2004 + A2:2006 / EN 55011:2007 + A2:2007EMC Generic emission:EMV Störfestigkeit: IEC 61326-1:2005 / EN 61326-1:2006 Industrie BereichEMC Generic immunity:Industrial environmentErgänzende Informationen: Batteriegerät! Bei Benutzung des Gerätes mit einem AC/DC-Adapter (Option ) ist nur Typ<strong>4411</strong>-Z001 (B9108WB) bei der der Erfüllung der Sicherheitsbestimmungen zugelassen.Additional Information: Battery device! When using with AC adaptor (optional), only B9108WB conforms to the safety regulation.Das Produkt wurde in einer typischen Konfiguration getestet.The product was tested in a typical configuration.Diese Konformitätserklärung betrifft alle nach Ausstellungsdatum ausgelieferten Produkte:This DoC applies to above-listed products placed on the EU market after:Gernsbach 09.07.2008 i.V. Alfred GroßmannDatum / dateQuality ManagerDieses Dokument ist entsprechend EN ISO/IEC 17050-1:2004 Abs. 6.1g ohne Unterschrift gültig / According EN ISO/IEC 17050 this document is valid without a signature.burster präzisionsmesstechnik gmbh & co kg . Talstr. 1-5 . D-76593 Gernsbach (Postfach 1432 D-76587 Gernsbach) Tel. 07224/645-0 . Fax 645-88www.burster.de ..www.burster.com info@burster.deSitz der Gesellschaft: HRA 530170 Mannheim . Komplementär: burster präzisionsmesstechnik Verwaltungs-GmbH . Sitz der Gesellschaft: Gernsbach HRB 530130 Mannheim...Geschäftsführer: Matthias <strong>Burster</strong> Prokurist: Edgar Miggler UST-Identnr.: DE 144 005 098 Steuernr. : 39454/10503Dresdner Bank AG Rastatt Kto. 06 307 073 00 BLZ 662 800 53. Volksbank Baden-Baden*Rastatt eG Kto. 302 082 00 BLZ 662 900 00Page 3


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®Table of contentsPage1. General .................................................................. 91.1 Application .............................................................. 91.2 Description .............................................................. 92. Safety Precaution ............................................... 113. Names and Function of Parts ............................ 134. Replacing Batteries ............................................ 195. Turning the Power ON/OFF ................................ 215.1. Switching ON to OFF ............................................ 226. SOURCE (generation) ......................................... 236.1. Connecting the output terminal ............................ 236.2. Generating DC current or DC voltage ................... 247. Measurement ...................................................... 277.1. Connecting procedure .......................................... 287.2. Measuring DC voltage and DC current ................. 298. Other Features .................................................... 318.1 Sweep function ..................................................... 318.2 Switching display resolution ................................. 348.3 Disabling automatic power OFF .......................... 349. Technical Data ..................................................... 3510. How to Use the Carrying Case .......................... 3711. Disposal ............................................................... 39Page 5


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 6


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®<strong>Voltage</strong>/<strong>Current</strong> <strong>Calibrator</strong>DIGISTANT ®Page 7


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 8


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®1. General1.1 ApplicationThe DIGISTANT ® model <strong>4411</strong> is capable as a small manageable checkingand calibration device for process technology in field.The other applications in the metrology include e.g. the calibration ofindication instruments, recorders, the checking of amplifiers, transducersand so on...The interests of the control technology were included particularly. Thedirect calibration data entry by UP/DOWN keys, the use of ramp functions,simplifier the management of measurement tasks.The function SINK can simulate a 2 wire transducer in a current loop.Both current and voltage can be generated or measured.1.2. DescriptionThe operation is made by UP/DOWN keys for each digit, measuringrange and measuring functions are activated by rotary switches and orslide switches.The DIGISTANT ® <strong>4411</strong> generates and measures voltages up to 30 V DCand currents up to 24 mA DC.The output level can be changed up and down between 4, 8, 12, 16 and20 mA signals by one touch for the “4 - 20 mA” output, and between1, 2, 3, 4 and 5 V DC signals for the “1 ... 5 V” output, for efficient calibrationwork.The “20 mA SINK” function absorbs a current adjustable from0.1 … 24 mA in a LOOP with a voltage from 5 … 28 V DC.The “ramp function” increases and decreases the output level to thepreset level at a constant rate for the selected sweep time (16 or 32seconds).Page 9


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 10


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®2. Safety PrecautionsWhen operating the instrument, strictly observe the precautionsbelow to ensure its correct and safe operation. If used other thanas instructed in this manual, burster präzisionsmesstechnik is notliable for any damage that may result.►The following safety symbols are used on the instrument and in themanual:Danger! Handle with care.This symbol indicated that the operator must refer to anexplanation in the Operation Manual in order to avoidrisk of injury or death of personnel or damage to the instrument.WARNINGIndicates a hazard that may result in the loss of life orserious injury of the user unless the described instructionis abided by.CAUTIONIndicates a hazard that may result in an injury to the userand/or physical damage to the product or other equipmentunless the described instruction is abided by.NOTEIndicates information that is essential for handling theinstrument or, should be noted in order to familiarizeyourself with the instruments operating procedures and/or functions.Page 11


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Damage to the instrument, personal injury or even death mayresult from electrical shock or other factors. To avoid this, followthe precautions below.WARNING► Use where gases may be presentDo not operate the instrument in a location where flammableor explosive gas/vapor present. It is extremelyhazardous to operate the instrument in such an atmosphere.► External connectionIf necessary to touch a circuit to make an externalconnection, turn off the power to that circuit, ensurethere is no voltage, then perform the connection.When replacing the batteries, disconnect the leadcables in advance.► DisassemblyDo not disassembly or remodel the instrument yourself.This needs to be done by our service personnel.Page 12


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®3. Names and Functions of PartsFront ViewSide ViewDIGISTANT <strong>4411</strong> Display unit AC adapterconnection jack POWER key OUTPUTON/OFF key Output valuesetting keys Range selectionrotary switch MEASURE(measurement)/SOURCE(generation)selection switch <strong>Voltage</strong> inputterminals <strong>Voltage</strong> output/current input andoutput terminalsRedBlackRear View(with the rear cover removed) DIP switches Battery holder Lead cablesPage 13


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®fMain seven segment, displays a measured value or anoutput value.gSub seven segment "SP" appears here when the sweepfunction (see section 8, "Other Features") is selected inSOURCE mode (signal generation). It also displays thelower two digits of a measured or an output value in thecalibration mode.hDisplays the unit of the range selected.iOFFIn SOURCE mode (signal generation), it lights up when theoutput is turned off or when the protective circuit isactivated.In MEASURE mode (measurement), it lights up only whenthe protective circuit is activated.jONIt lights up when the output is turned on in SOURCE mode(signal generation). OUTPUT value setting keysSets an output value SOURCE mode (signal generation).The [▲/▼] keys are provided under the lower four digits, whosevalues are increased or decreased in instruments of 1. Carry ofthe digit is applied to increasing the value (pressing in [▲] keywhen it is 9. Borrow of the digit is applied to decreasing thevalue (pressing the [▼] key when it is 0.For the 4-20 mA and 1 - 5 V, see subsection 6.2, "GeneratingDC current or DC voltage".Page 15


DIGISTANT ® <strong>Model</strong> <strong>4411</strong> POWER keyTurns on/off the power supply. For more information, see section5, "Turning the Power On/Off". OUTPUT ON/OFF keyTurns the output on/off in SOURCE mode (signal generation). InMEASURE mode (measurement), it returns measurement operationto normal status after the protective circuit is activated. Range selection rotary switchSelects a range for SOURCE mode (signal generation) orMEASURE mode (measurement). Note that the 4 - 20 mA and1 - 5 V ranges are step output ranges in signal generation andare the same as the 20 mA and 10 V ranges in measurement,respectively. For invalid ranges, terminal-to-terminal connectionbecomes open, causing "-nC-" and "OFF" to appear on thedisplay unit. MEASURE (measurement) /SOURCE (generation) selection switchSelects MEASURE (measurement) or SOURCE (generation). <strong>Voltage</strong> input terminalsUsed to measure a voltage range. <strong>Voltage</strong> output/current input andoutput terminalsUsed for SOURCE mode (signal generation) in the voltagerange, and for MEASURE mode (measurement) and SOURCEmode (signal generation) in the current range.Page 16


<strong>Model</strong> <strong>4411</strong> DIGISTANT ® AC adapter connection jackUsed to connect an AC adapter (optional). DIP switchesSee section 8 "Other Features". Battery holderContains four AA-size batteries.See section 4, "Replacing the Batteries". Lead cables for measurement or generationUsed to connect the instrument to the device undermeasurement/generation.Notice regarding the manual The information contained in this Operation Manual is subjectto change without notice. Every effort has been made to ensure that the informationcontained herein is accurate. However, should anyconcerns, errors, or omissions come to your attention, orif you have any comments, please contact us.Page 17


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 18


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®4. Replacing BatteriesWhen the + - mark is blinking on the display unit, the batteriesare exhausted. Replace them according to the following procedure: Check that the power is turned off (disconnect the lead cables). Slide off the cover at the back of the instrument. Replace all four batteries with new ones. Insert them according tothe polarity directions shown inside the holder. After replacing the batteries, return the cover to the original position.Page 19


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>►Connecting the AC power (optional)Before connecting the AC power strictly observe thefollowing warnings to avoid electrical shock or damageto the instrument.WARNING► Before connecting the power cord, check that the supplyvoltage matches the rated voltage of the instrument.► Before connecting the power cord, check that theinstruments power key is OFF.► Do not use any AC adapter other than the dedicatedAC adapter.Connection procedure Check that the [POWER] key of the instrument is off. Connect the AC adapter (optional) to the instruments ACadapter connection jack. (Note that unless the AC adapteris connected to the power outlet, the power cannot beturned on).Page 20


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®5. Turning the Power ON/OFF► Operating the POWER keyWhen the instruments power is off, pressing and holding the[POWER] key for more than 1 second causes the power to beturned on. Pressing the key again causes it to be turned off.When the power is turned on, the instrument starts self-test anddisplay "CA11". Then the features selected using the range selectionrotary switch and the MEASURE/SOURCE selection switchstart functioning.► For battery-driven operations, disconnect the AC adapterfrom the instrument.► Automatic Power OFFIn the factory setting, all indications start blinking if the instrumenthas not been operated for about 9.5 minutes. Then, if the instrumentis not operated for another 30 seconds, it automaticallyturns off.To disable this automatic power off feature, see section 8 "OtherFeatures".If you wish to keep the instrument turned on after the indicationsstart blinking, press the [POWER] key (or any other key). Thiscauses the blinking to return to normal lighting, without changingthe previous settings.Page 21


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>5.1 Switching ON to OFFThe Lo terminal of <strong>4411</strong> will react with cut-off of the internal circuitthrough a switch, when OUTPUT is turned off.Page 22


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®6. SOURCE (generation)6.1 Connecting the output terminal Insert the plugs of the supplied lead cables into the output terminalsof the instrument. Connect the clips on the other ends of the cables to the inputterminals of the device under generation.OUTPUT V/mAHiLoHi– +CAUTION►►Do not apply any voltage to the output terminals except for the20 mA SINK range. If voltage is applied mistakenly, the internalcircuit may be damaged.As the instrument is calibrated without the voltage drop of thelead cables, error due to the resistance of the lead cables (approx.0.1 Ω for go an return) must be considered for load currentmeasurement.Page 23


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>6.2 Generating DC current or DC voltageThe instrument generates voltage or current at a specified value throughthe output terminals. In valid ranges, terminal-to-terminal connectionopens, and the display unit indicates "-nC-" and "OFF".►Normal setting: Switch the MEASURE/SOURCE selection switch to "SOURCE"(generation). This causes the display unit to show "SOURCE"and "OFF". Select the range to be generated using the range selectionrotary switch. The display unit shows an initial value and unitfor each range. Press the [▲] / [▼] keys under the lower four digits to set anoutput value (5ft digit: changes by carry and borrow of thevalue from the 4th digit). Press the [OUTPUT ON/OFF] key to output a signal. The displayunit shows "ON".To stop the output, press the [OUTPUT ON/OFF] keys again.This causes "OFF" to appear on the display unit and theterminals to open.Page 24


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®►►►►Allowable loadThe instrument should be used according to the maximum output andthe maximum load range (see section 10, Technical Data").<strong>Current</strong> SINK (absorb) is only possible in the 20 mA SINK range.1 - 5 V step output rangeIn the 1 - 5 V range, the [▲] / [▼] keys for the most significant digit canbe used to step up or down by 1 V. For the lower three digits, these keysare used to increase or decrease the value of each digit in incrementsof 1.4 - 20 mA step output rangeUse the range selection rotary switch to select 4 - 20 mA5-digit resolution, set dip switch 3 to "OFF".Use the [▲] / [▼] keys under the upper one digit to change the value inincrements of 4 mA: 4 ←→ 8 ←→ 12 ←→ 16 ←→ 20 mA.For the lower three digits, these keys are used to increase or decreasethe value of each digit in increments of 1.4-digit resolution, set dip switch 3 to "ON".Use the [▲] / [▼] keys under the upper two digit to change the value inincrements of 4 mA: 4 ←→ 8 ←→ 12 ←→ 16 ←→ 20 mA.For the lower two digits, these keys are used to increase or decreasethe value of each digit in increments of 1.24 V (20 mA) rangeIn the 24 V output (20 mA measurement) range, the instrument can beused as a source of constant voltage (24 V) to conduct a loop test of atransmitter.Select "24 V (20 mA)" and SOURCE using the respective switches.Press the "OUTPUT ON/OFF" key to set the output to "ON" (for 24 V DCoutput). The display unit shows the current measurement and "SOURCE"(for 20 mA measurement).Page 25


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>► 20 mA SINK rangeThe instrument can absorb (SINK) a current with a specified value sentfrom an external voltage generator (distributor) to the "Hi" terminal ofthe instrument. As the instrument serves as a two-wire transmitter, itcan be used for a loop test. Select "20 mA SINK and SOURCE using therespective switches. Press the "OUTPUT ON/OFF" key to set the outputto "ON". The display unit shows the set value (SINK) and "SOURCE".Uses an external power supply for 20 mA SINK in the range of 5 - 28 V.The polarity of applied voltage should be as shown in the figures below.Be careful not to apply voltage in the opposite direction.►►Output limiterIf a load current in the voltage range or a load voltage in the currentrange exceeds the maximum value specified in the specifications, theprotective limiter is activated, turning off the output signal. To recoverfrom this condition, correct the load to normal status and press the"OUTPUT ON/OFF" key to turn on output.Sweep functionThis function increases or decreases the output level to the preset levelas a constant rate by setting DIP switch 1 to ON. For more information,see subcategory 8.1 "Sweep Function".Page 26


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®7. MeasurementWARNING► When connecting the device under measurement, turn off the powerof the device. Connecting/disconnecting the lead cables for measurementwithout turning off the power of the device under measurementmay be extremely dangerous.► Special care should be taken to avoid connecting a current circuitto the voltage input terminals or a voltage circuit to the current inputterminals. Inadvertent connection may not only cause damage to thecircuit or device under measurement and this instrument, but mayalso be dangerous to personnel.► Always use the lead cables supplied with the instrument.► Maximum allowable voltage between all input/output terminals andground is 30 V DC. Any voltage exceeding this level may not onlydamage the instrument, but also cause injury to personnel. Neverattempt to apply such voltage.Page 27


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>CAUTION► Do not apply voltage exceeding the maximum allowable input voltageto the voltage input terminals. This may damage the instrument.Maximum input voltage: 30 V DC► Do not apply current exceeding the maximum allowable input to thecurrent input terminals. If current exceeding this level is applied, buildinfuses for protecting the current input circuit will burn out. If a fuseburns out, contact us for repair.Maximum input current: 24 mA7.1 Connecting procedure Insert the plug of the supplied lead cables into the terminals ofthe <strong>4411</strong> as shown in figure "a" below for voltage or in figure"b" below for current. Connect the clips on the other ends to the input terminals ofthe device under measurement.a Connecting of thevoltage input terminalsb Connecting of thecurrent input terminalsINPUT VINPUT mAHi Lo Hi Hi Lo Hi+ –– +Page 28


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®7.2 Measuring DC voltage and DC current Switch the MEASURE/SOURCE selection switch to MEASURE(measurement). MEASURE lights up on the display unit. Select the range to be measured using the range selection rotaryswitch.► The display unit shows the measurement results and unit.Indication of the measurement results will be updated approximatelyevery second.► If an input signal is over range or not measurable, the displayunit shows "____".► For invalid ranges, terminal-to-terminal connection becomesopen, causing the display unit to show "-nC-" and "OFF".►Input limiterIf voltage is applied to the current input terminals or excessive currentis applied to them, the internal protective circuit is activated, openingthe terminal-to-terminal connection and causing the display unit toshow "OFF".In this case, check that the connection is correct and press the "IN-PUT ON" key (= "OUTPUT ON/OFF" key). This causes the OFF indicationto disappear and measurement to resume.Page 29


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 30


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®8. Other FeaturesThe following features are available depending on the settings of theDIP switches, which can be accessed by removing the cover from theback of the instrument.DIP switchesON1 2 3 41: Sweep function2: Sweep time (ON: 32 sec. OFF: 16 sec.)3: <strong>Current</strong> display resolutionOFF: 5 digitsON: 4 digits4: Disable automatic power off8.1 Sweep functionBy setting DIP switch 1 to ON, the output level can be increased or decreasedto the preset level at a constant rate.This function cannot be used in the 24 V (20 mA) range.Output status transition diagramSet valueHigh impedanceLower range limitSweep status(A) (B) (C) (D)(E) (F) (A)Operation ofOUTPUT ON/OFF key↑ (↑) ↑(↑)Page 31


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Operation procedure Turn on DIP switch 1. Use DIP switch 2 to select sweeping time (16 or 32 seconds). Set the MEASURE/SOURCE selection switch to SOURCE. Thedisplay unit shows "SOURCE", "OFF", "SP", and lower limit (initialvalue). Use the [▲] / [▼] keys to preset the upper limit for output in thesame way as for normal signal generation (status A above). Press the "OUTPUT ON/OFF" key The display unit shows "ON" and a lower limit (initial value) for 2seconds, and the instrument outputs that value. Then the reading on the display unit and the output level automaticallyincrease to the upper limit set in step at the constantsweep time rate set in step .The display unit shows the "ON" light blinking.In this case, if you press the "OUTPUT ON/OFF" key again, thereading and the output level will start decreasing from that valueat a constant rate. The output status changes to status E (shownby dotted-lines in the figure).Page 32


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®When the reading and the output level reach the set values,they remain at these levels until the"OUTPUT ON/OFF" key ispressed. The display unit´s "ON" light comes on.In this case, if you press the"OUTPUT ON/OFF" key, the readingand the output level start decreasing from that value at aconstant rate. The output status changes to status E.Then the reading and the output level automatically decreaseto the set lower limits at the constant sweep time rate set instep . The display unit shows the "OFF" light blinking.In this case, if you press the "OUTPUT ON/OFF" key again, thereading and the output level start increasing from that value ata constant rate.The output status changes to status C (shown by dotted-linesin the figure).Then the reading and the output level reach the lower limits andremain at these levels for 3 seconds. The display unit´s "ON"light comes on. Return to step .Cancelling the sweep function: Open the cover at the back of the instrument and set DIP switch1 to off.Page 33


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>8.2 Switching display resolutionBy changing the setting of DIP switch 3, the display resolution (5 digit/4digit) can be selected.The available ranges are as follows:SOURCE: 20 mA, 4 - 20 mA, 24 V (20 mA), 20 mA SINKMEASURE: 20 mA, (4 - 20 mA)Example:4 - 20 mA ranges5 digit range 4 digit range8.3 Disabling automatic power OFFAccording to the factory setting of the instrument, it automatically turnsitself off if not operated for about 10 minutes. By setting 4 to "ON", thisfunction can be disabled.However, when the instrument is battery-driven, it is generally recommendedthat this switch be set to "OFF" in order to prevent the batteriesfrom being exhausted.Page 34


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®9. Technical Data►SOURCE:Accuracy: ±(% of set value + µV, mV, or µA) at 23 ± 5°CRange Range of Accuracy Setting RemarksSelection Generated Signal Resolution30 V0 to 30.00 V0.05 % + 20 mV10 mV Maximum output current: 1 mA10 V1 V0 to 11.000 V1/2/3/4/5 V0 to 1.1000 V0.05 % + 2 mV0.05 % + 0.2 mV1 mV1 Vstep0.1 mVMaximum output current: 10 mA*When the load is 1 kΩ or greater andthe error of the lead cables is alsoexcluded.100 mV0 to 110.00 mV0.05 % + 50 µV10 µV20 mA*4 - 20 mA*0 to 24.000 mA4/8/12/16/20 mA0.05 % + 4 µA1 µA4 mAstepMaximum load: 12 V24 V (20 mA)*24 V± 10 %-Maximum current: 22 mA20 mA SINK*0.1 to 24.000 mA0.1 % + 4 µA1 µAExternal power supply: 5 to 28 VTemperature coefficient: 1/10 of accuracy/°C, but (0.005 % + 10 µV)/°C in the 100 mVrange.* The resolution can be changed between 5 digits and 4 digits using a DIP switch.►MEASURE:Accuracy: ±(% of reading + least significant digits), at 23 ± 5°CRange Indication Accuracy Resolution RemarksSelection30 V0 to ± 30.00 V0.05 % + 2 dig.10 mV10 V1 V0 to ± 11.000 V0 to ± 1.1000 V0.05 % + 2 dig.0.05 % + 2 dig.1 mV0.1 mVInput impedance: approx. 1 MΩ100 mV0 to ± 110.00 mV0.05 % + 7 dig.10 µV20 mA*0 to ± 24.00 mA 0.05 % + 4 dig.1 µA Input impedance: 45 ΩTemperature coefficient: 1/10 of accuracy/°C, but (0.005 % + 10 µV)/°C in the 110 mVrange.* The resolution can be changed between 5 digits and 4 digits using a DIP switch.Power supply:Battery life:Four 1.5 V alkaline batteries (LR6, AA-size)or dedicated AC adapter (sold seperately)Approximately 50 hours for 5 V DC output with a load of 10 kΩor greater (when running on alkaline batteries).Approximately 25 hours for 20 mA output with load of 5 V(when running on alkaline batteries).Page 35


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Automatic power off:after a period of approx. 10 minuteswith no operationsGeneration signal level setting:by four sets of "up" and "down" keysResponse of generator:approx. 1 second(between the time for an output signal to start changingand the time when it enters the accuracy range)Loading conditions: < 0.1 µF (DC V)Measured-value indication updating intervals:Display:Maximum allowable applied voltage:Operating temperature and humidity range:Storage temperature and humidity range:Dimension ( H x W x D):Weight:Safety standards:EMC standards:Immunity:Page 36approx. 1 second7 segment LCD30 V DC betweeneach terminal and ground0 ... 50 °C,to 20 ... 80 % R.H.(no condensation)-20 ... 50 °C,90 % R.H. or less(no condensation)approx. 192 x 92 x 42 mm(excluding protrusions))approx. 440 gEN61010-1 (only AC adapter <strong>4411</strong>-Z001)EN61326 class BEN55011 group 1, class BEN61000-3-2EN61000-3-3EN61326Performance criterion under immunity testenvironments: B(self-returnable performance deterioration)Test conditions of EMC and immunity standards:AC adapter and lead cables <strong>4411</strong>-Z001Accessories:connecting cable (1 pair), 4 alkaline batteries,carrying case, operation manualOptional accessories: dedicated AC adapter <strong>4411</strong>-Z001


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®10. How to Use the Carrying CaseThe carrying case may be used as follows(1) Undo the fasteners onthe top and sides ofthe case cover to openit.(2) With the fastener onthe logo-side of thecase centered, lift thecover and pivot it tothe side and under thecase itself.(3) Re-do the fasteners atthe top and sides ofthe cover.Other features of the carrying case and instructions► The device can be mounted with the terminals connected to thedevice interface cables in the housing on the back of the trunklid.► The strap makes it possible to use the device or stored by it ona hook, rod, etc. depends.Page 37


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 38


<strong>Model</strong> <strong>4411</strong> DIGISTANT ®11. DisposalBattery disposalAs an end user, you are required by law (batteryordinance) to return all used batteries and rechargeablebatteries; the disposal through householdwaste is prohibited. By buying the herein describeddevice you are concerned by this law. Please disposeof your batteries and rechargeable batteriescorrectly. Hand them to waste disposal sites eitherat your premises or at our company or at any placewhere batteries/rechargeable batteries are sold.Equipment DisposalPlease fulfill your legal obligations and dispose ofunserviceable equipment in accordance with applicablelegal requirements. Thus you contribute toenvironmental protection.Page 39


DIGISTANT ® <strong>Model</strong> <strong>4411</strong>Page 40