<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-I8.14 Maintenance LogThe Maintenance Log is used to know the last time the Heater, Temperature Probes or Motorwere calibrated. As well, the Maintenance Log will tell the user or service person the amountof run time associated with the Belts, Motor, and Heater. The Maintenance Log is forreference information only; the log should only be cleared or changed by a service technicianafter service has been conducted or one of these components has been replaced.1. With the Main screen displayed, press [PREFERENCES] to display the SystemOptions screen.2. Press [3] to select "Validation".3. Press [3] to display the “Maintenance Log” screen.The Maintenance Log may be printed.Only belt, motor, and heater time may be reset.8.15 Tablet CalibrationTablet calibration of the <strong>SR8</strong>-Plus is accomplished by following the guidelines listed in theUSP-NF under Dissolution. This test should be performed after the installation is complete.It is normal to do tablet calibration every 6 months. Use the following checklist whencalibration is due.Check ListRecommended ToleranceTemperature±0.5°CSpeed