<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-I8.9 Guide RodsThe drive head should move smoothly whenever raising or lowering it. Lubricate the guiderods with silicone grease whenever smooth movement is not observed.8.10 Spindle BearingsFree rotation of the spindle bearings should be checked every 6 months.To check the spindle bearings:1. Turn the motor off by placing the system in stand by or by turning the entire system off.2. Disengage the clutch by lifting it off the belt pulley and turning 90º. Turn each spindle byhand, as shown in Figure 8-2 and check for irregular rotation or sticking.Figure 8-2: Disengaging the Clutch3. Each spindle should turn freely with no binding or tight spots. If there is a problem,contact Hanson Research.NOTE: Only use lithium grease to lubricate the spindle bearings.4. After checking each spindle bearing, engage the clutch and turn the system back on.Section 8 Maintenance and Calibration – 57 –