<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-ITo calibrate the temperature probe(s):1. Display Main screen, press [PREFERENCES] to display the System Optionsscreen.1. Press [4] to display the Calibration menu.2. Press [2] to select “Temperature Probe Calibration”.3. After reading the warning screen, Press [ENTER].4. Press [SET TEMPERATURE] to display Temperature Calibration screen. (seeFigure 8-1) The bath will heat to about 32°C (approximately 30 minutes).Figure 8-1: Temperature Calibration Screen5. Place a certified thermometer and all temperature probe(s) adjacent to each other in thewaterbath. When the reading has stabilized (approximately 30 minutes), inputthermometer temperature reading and press [ENTER].6. The bath will then go to 37°C (approximately 30 minutes). When the reading hasstabilized, input thermometer temperature reading and press [ENTER].7. Check temperature probe accuracy. (see Section 8.5—Temperature Probe AccuracyCheck)NOTE: If the temperature probes have been calibrated and still do not match thethermometer reading within 0.1°C, see Section 8.6—Temperature Probe OffsetAdjust.Section 8 Maintenance and Calibration – 55 –