<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-I6.5 During the Test1. The <strong>SR8</strong>-Plus will count down to each sampling time point. The display will showelapsed time, the bath temperature, and spindle speed.2. The <strong>SR8</strong>-Plus will also send important information to the validation printer for a hardcopytest report of the process if a printer is available and programmed to do so.6.6 After the TestWhen the test is complete the <strong>SR8</strong>-Plus will print (if a printer is available) the performancesummary which includes: Minimum, Maximum and Average for the speed andtemperature(s).The <strong>SR8</strong>-Plus will automatically stop the spindles.The heater will shutdown if programmed to do so.Section 6 Running A Test – 49 –