<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-I5. Press the up arrow to return the sampling probes to the home position.NOTE: The probe will only move if a stroke value has been entered into the protocol. Alwaysset the stroke to 45mm unless there is a requirement to stroke a different amount.4.14 Adjusting Digital Temperature ProbesThe tip of the Digital Temperature Probes should be adjusted to the tip of the sampling probeor its filter.To adjust Digital Temperature Probes:1. Make sure that the E-Probes are at the top of their stroke. If you are not sure, homeprobes. (see Section 5.17--System Configuration Parameters: E-Probe Setup)2. Have the <strong>SR8</strong>-Plus mechanical settings adjusted and ready to start the test.3. The Digital Temperature Probe is designed to clip onto the sampling probe. Move theDigital Temperature Probes up or down so that the tip of the Digital Temperature Probeis at the same level as the tip of the sampling probe or sampling filter.Section 4 Installation – 23 –