<strong>SR8</strong>-Plus Operation <strong>Manual</strong>73-100-810-I3. Verify that the heater is working properly. Press [SET TEMPERATURE].Use the [↑] and [↓] arrow keys to highlight 37.0, and press [ENTER]. The “Heater On”status light should illuminate.4. The heater should raise the bath temperature to close to 37°C within 30 minutes andequilibrate at 37° in one hour. The temperature can be monitored on the status screen.5. Run a diagnostics test and verify that all the tests pass.NOTE: If any test fails or if the bath does not heat, contact Hanson Research TechnicalSupport at (800) 821-8165 or (818) 882-7266.4.10 Setting Adjustable Stops1. Raise the drive head to its highest position.2. Use a hex wrench (supplied in the supply kit) to loosen the Adjustable Stops. Set thestops to where there is 38mm (1.5”) of distance between the Adjustable Stop and thesupport that is on top of the base. The width of all height spacers is 38mm; an easy wayto set this distance is to use the height spacer on its side and place it between the stopand the support. Set one of the Adjustable Stops this way. Tighten this stop to lock inposition. Lower drive head so that it touches this stop and then lift the secondAdjustable Stop so that it touches the drive and lock into place.Figure 4-14: Setting Adjustable StopsSection 4 Installation – 20 –