4"12-Z 1 - part - usaid
4"12-Z 1 - part - usaid
4"12-Z 1 - part - usaid
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
14<br />
The 1750<br />
Proximate<br />
Analyzer<br />
A rapid data processing<br />
system for calculating and<br />
reporting the proximate<br />
analysis of coal.<br />
RECOIvfMENDED FOR<br />
Assembling data, calculating and reporting the<br />
proximate analysIs of coal, coke and related fuel<br />
samples as tested In coal mine, coal preparation<br />
plant, coal burning utility, commercial Inspection and<br />
other Industrial laboratories where coal samples are<br />
tested daJly on a routine basIs.<br />
A FULLY INTEGRATED<br />
DATA PROCESSING SYSTEM<br />
The proximate analysIs of a coal sample IS a tedious<br />
procedure involving several separate welghmgs and<br />
producing considerable data as needed to calculate:<br />
Air-dry moisture Volatile matter<br />
Total mOisture Fixed carbon<br />
Ash<br />
There IS no way to avoid the indIVIdual welghlngs<br />
and heatlngs Involved in these separate<br />
determinations, but the data entry, data processing<br />
and reporting functions can be speeded and<br />
simplified considerably uSing the convenient<br />
microprocessor system provided by a Parr 1750<br />
Proximate Analyzer.<br />
Operating in conjunction with a mOisture oven.<br />
furnace and analytical balance already on hand In a<br />
user's laboratory. the 1750 Proximate Analyzer Will<br />
• Accept and retain all individual weighings<br />
obtained dUring the series of steps required to<br />
determine air-dry mOisture, total mOIsture, volatile<br />
matter, fixed carbon and ash.<br />
• Organize and store all test data for up to 100<br />
samples, accepting the data as It becomes available<br />
from a digital balance Samples can be weighed and<br />
treated In any order with the Analyzer accepting the<br />
data as generated. organizing and storIng It In proper<br />
sequence under operator assigned sample numbers<br />
• Collect and store sample weights for<br />
calorimetric determinations. If a 1750 Proximate<br />
Analyzer IS used With a <strong>12</strong>61 CalOrimeter (or With a<br />
<strong>12</strong>41 CalOrimeter eqUipped With a 1720 or 1730<br />
Controller) the 1750 Analyzer Will provide all of the<br />
funcoons of a 1741 Balance Interface, and a separate<br />
Balance Interface Will not be reqUired. Each welghmg<br />
Will be shown on an LED display before It IS entered<br />
Into memory, With the Analyzer promptmg the<br />
operator to Identify each entry With a valid sample<br />
number. Weights can also be entered from a<br />
keyboard The Analyzer Will collect and hold all<br />
sample weights for calOrimetric determination untIl<br />
called up by the calOrimeter.<br />
• Collect and store sulfur values from a 1760 Sulfur<br />
Analyzer to be Included in a consolidated proximate<br />
analystS report.<br />
• Calculate air-dry moisture, total moisture,<br />
volatile matter, fixed carbon and ash for each<br />
sample on any of four reference bases:<br />
As-received (AR) Dry (DRY)<br />
As-determmed (AD) Dry, ash-free (OAF)<br />
• Convert calorific and sulfur values to any of the<br />
mOisture reference bases listed above.<br />
• Produce a consolidated proximate analyses<br />
report for each valId sample (includmg calOrifIc value<br />
and sulfur) by feedmg all collected reports to any of<br />
the folloWing:<br />
A Parr 1755, 40-column printer or an 80-column<br />
commercial printer.<br />
A <strong>12</strong>61 Calonmeter WIth a Parr 1755 Printer<br />
A central computer.<br />
A laboratory or personal computer connected on a<br />
Parr Smart Link.<br />
• Produce full or <strong>part</strong>ial reports at any time for an<br />
indIVIdual sample or for a full range of tests Reports<br />
can Include all supporting raw data or can be<br />
condensed to show only final values.<br />
("')<br />
) -