12.07.2015 Views

Flexi-Q DV Auto-injector Case Study - DALI Medical Devices

Flexi-Q DV Auto-injector Case Study - DALI Medical Devices

Flexi-Q DV Auto-injector Case Study - DALI Medical Devices

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Study</strong> Conclusions• The study results establish that the users can perform all criticaltasks without a pattern of preventable use error after training decay• All of the participants who read and followed the IFU were able touse the <strong>injector</strong> successfully• A small residual risk remains that certain patients may ignore thetraining and instructions which may result in mis-dosing.This risk is one that cannot be eliminated through design orlabeling changes and is a risk that is residual for almost allinjectable drugs, regardless of delivery system (even in prefilledsystems)• Additional information submission to FDA (Jan’ 2012)July2009Feb’2010Nov’2010May2011July2011Jan’201227

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!