Flexi-Q DV Auto-injector Case Study - DALI Medical Devices

Flexi-Q DV Auto-injector Case Study - DALI Medical Devices Flexi-Q DV Auto-injector Case Study - DALI Medical Devices

12.07.2015 Views

Device and Changes Validation• Simulated Clinical Use (Summative) Study• Performed based on the FDA CDRH Guidance:“Medical Devices With Sharps Injury PreventionFeatures”14

Study Methodology - Brief1. Devices tested:500 units of Flexi-Q DV auto-injectors2. 30 total naïve participants recruited:– 5 participants being mild to moderate RA/MS patients– Male and female aged 18-70 years old3. Each participant operated 15-17 Flexi-Q DV’s:– Injecting into foam-pad positioned:on her/his thigh (~½);on her/his abdomen (~½);15

Device and Changes Validation• Simulated Clinical Use (Summative) <strong>Study</strong>• Performed based on the FDA CDRH Guidance:“<strong>Medical</strong> <strong>Devices</strong> With Sharps Injury PreventionFeatures”14