Page 33<strong>Transcend</strong> Sleep Apnea Therapy <strong>User</strong> <strong>Manual</strong>Alert codesAlerts are similar to faults but they do not reset the processor when they occur (i.e., thedevice continues to deliver therapy pressure).Device LEDOffFaultLEDFlashes15 timesErrorCannot regulateCommentsDevice repeatedly detected the monitored pressureoutside of the therapy target range. The deviceattempts to continue to deliver therapy. The alert is anindication that filters and the mask should be inspectedfor cleanliness and proper fit.Filter Media change reminderEvery six months, the green LED flashes as a reminder that the Filter Media should bechanged. The LED only flashes when the <strong>CPAP</strong> is in Standby Mode. The <strong>CPAP</strong> willcontinue to function normally. The reminder may be acknowledged in a manner similar toacknowledging a fault (hold down the power button until the green LED stops flashing).Acknowledging the reminder resets the reminder timer, that is it will flash again in sixmonths. The reminder should be acknowledged in conjunction with changing the filter.