EXTENDED REGISTER GROUPS<br />
OptAD, analog to digital converter.<br />
OPTAD, ANALOG TO DIGITAL CONVERTER.<br />
Group Group<br />
No.<br />
Description<br />
OptAD 22 Optional analog to digital conversion.<br />
GENERAL<br />
The DMC equipped with this option has an analog input named M1. This channel<br />
has an individual gain setting. The gain setting is controlled with the<br />
OptAD.GainM1 member. The input is intended for a strain gauges transducer i.e. a<br />
load cell or a torque transducer. The connector also incorporates a balanced<br />
supply voltage <strong>of</strong> +/- 5 VDC.<br />
Channel Measuring range Purpose<br />
M1 ±50 mVDC Strain gauge<br />
transducer<br />
CONVERSION RESULOTION<br />
Extern measurement<br />
bridge is required.<br />
The AD resolution is 14 bit, and the result is sign extended to utilize the 32 bit PL<br />
register set.<br />
CALIBRATION<br />
The M1 channel has two resistances, 390 kΩ and 47kΩ, that can be used to<br />
calibrate an externally provided measurement bridge. When the value in the table<br />
is written to OptAD.Cal then the selected calibration resistance will be connected in<br />
parallel with one <strong>of</strong> the four legs in the Measurement Bridge.<br />
Value Channel Resulting calibration resistance.<br />
0 none Disconnect all calibration resistances.<br />
8 M1 390 kΩ<br />
9 M1 47 kΩ<br />
EXAMPLE USAGE<br />
; On startup.<br />
OptAD.Cal = 0 ; disconnect all calibration resistances.<br />
.. <br />
; Start calibration<br />
OptAD.Cal = 8 ; M1, calibration resistance = 390 kOhm<br />
R200 = OptAD.M2 ; Conversion is done within 200 us.<br />
.. <br />
OptAD.Cal = 0 ; disconnect all calibration resistances.<br />
<br />
; Start measuring<br />
R100 = OptAD.M1 ;and the result is stored in R100.<br />
; The calibration value in R200 is application dependent and<br />
; bridge dependent, and it can be used in several different<br />
ways.<br />
;<br />
; 1. Value may indicate what type <strong>of</strong> measurement<br />
User's Manual 5.1 Inmotion Technologies AB<br />
Doc. No.9032 0027 01 (B), Rev. 11.07.2001<br />
135