Without Temperature probe (no ATC)If no temperature probe is used, the meter compensates for pH response based on anew calibrated temperature value manually set by you or at 25.0 °C (factory default).1. Press to enter into Temperature mode until “°C” shows in LCD.2. Compare the displayed value to NIST certified thermometer or a thermometerknown to be accurate (dipped into a constant temperature bath).3. Press to enter temperature calibration mode. The LCD shows “CA”momentarily and displayed reading flashes. Note that this displayed value shouldeither be 25.0 °C or the last set temperature value.4. Press (for pH 5/6) until the displays shows the desired temperature. You canset any value from 0 to 100 °C.5. To cancel or abort this operation, press . Note no new value will be storedinto the meter‘s non-volatile memory. To confirm calibration, press . The LCDdisplays “CO” momentarily, and the meter reverts to measurement mode.Thermo Electron Corporation<strong>Russell</strong> <strong>RL060P</strong>13