2. Press [ZERO] to bypass Calibration Mode.3. Press [ZERO] to exit without saving changes.4. Press [TARE] to undo changes.5. Press [TARE] to complete exit.4.4.2 Model 355 I.S.To exit the Setup Mode and save changes1. Press [ZERO] to begin exiting Setup Mode.2. Press [CLR] to bypass Calibration Mode.3. Press [ENTER] to save setup changes.4. Press [ENTER] to complete exit.To exit the Setup Mode from the view-only mode1. Press [ZERO] to begin exiting Setup Mode.2. Press [ENTER] to complete exit.To exit the Setup Mode without saving changes1. Press [ZERO] to begin exiting Setup Mode.2. Press [CLR] to bypass Calibration Mode.3. Press [CLR] to exit without saving changes.4. Press [ENTER] to undo changes.5. Press [ENTER] to complete exit.4.5 Factory DefaultParameter 65001 and 65002 are available to return the Model 350/355 I.S. to factorysettings. Parameter 65001 will reset parameters including the calibration, whileparameter 65002 resets all parameters except the calibration.4.5.1 Return to factory default (350 I.S.):Access the setup mode. See page 47 for details.1. Key in [6] [5] [0] [0] [1] or [6] [5] [0] [0] [2] [SELECT]2. Press [TARE] once.3. Press [TARE] once.4. Press [ZERO] [ZERO] [TARE] [TARE]52Model 350 I.S. and Model 355 I.S. <strong>Service</strong> <strong>Manual</strong>