4. Attach the voltmeter - (black) lead to pin 5 (ISOLATED GND) of the AnalogOutput connector.5. Key in [6] [2] [0] [0] [3] [SELECT].6. Press [ENTER] to set the output to 0%.7. Press [ENTER] to increase the output to 25%.8. Press [ENTER] to increase the output to 50%.9. Press [ENTER] to increase the output to 75%.10. Press [ENTER] to increase the output to 100%.To test the 4-20mA output mode:Voltmeter readings are based on the use of a 500 ohm precision resistor.Caution! Do not exceed 500 ohms.1. Enter the Setup Mode (see Setup Mode on page 47).2. Attach the precision resistor to pin 5 and pin 6.3. Attach the voltmeter + (red) lead to pin 6 (4-20 mA) of the analog outputconnector.4. Attach the voltmeter - (black) lead to pin 5 (ISOLATED GND) of the AnalogOutput connector.5. Key in [6] [2] [0] [0] [4] [SELECT].6. Press [ENTER] to set the output to 0%.7. Press [ENTER] to increase the output to 25%.8. Press [ENTER] to increase the output to 50%.9. Press [ENTER] to increase the output to 75%.10. Press [ENTER] to increase the output to 100%.46Model 350 I.S. and Model 355 I.S. <strong>Service</strong> <strong>Manual</strong>