Only Zero Calibration (355 I.S.):1. From the Weigh Mode, press [1] [0] [0] [SELECT].2. Press [5] [4] [3] [2] ]1] [ENTER].3. Press [SELECT] [SELECT] [SELECT].4. Remove any load on the scale.5. Press [ENTER] to establish zero.6. Pause for motion delay.7. Press [ENTER] to accept calibration.8. Press [ENTER] to save calibration.9. Press [ENTER] to exit calibration.5.2.6 Reset CalibrationCal Reset may be necessary when an over-load or under-load condition exists,preventing the completion of the calibration process. Calibration Reset adjusts the zeroand gain factors of the A/D amplifier to factory default values for maximum sensitivity.After performing a calibration reset, a complete recalibration is required. The effects ofa calibration reset do not take effect until the Model 350/355 I.S. is recalibrated andcalibration information has been saved.If Code 02 (under-load) or Code 03 (over-load) is displayed during calibration, press[ZERO] or [CLR] to perform a calibration reset.Reset Calibration Gain Factors (350 I.S.):1. Press [ZERO] + [SELECT].2. Press [ZERO] [PRINT] [UNITS] [TARE].3. Remove any load on the scale.4. Press [TARE] to establish zero.5. Place a test weight on scale.6. Press [ZERO].7. Remove any load on the scale.8. Press [TARE] to establish zero.9. Pause for motion delay.10. Place a test weight on scale.11. Enter the test weight value with the [PRINT] and [UNITS] keys.12. Press [TARE] to establish span.122Model 350 I.S. and Model 355 I.S. Service Manual
13. Pause for motion delay.14. Press [TARE] to accept calibration.15. Press [TARE] to save calibration.16. Press [TARE] to exit calibration.17. Remove the calibration weight.Reset Calibration Gain Factors (355 I.S.):1. Press [1] [0] [0] [SELECT].2. Press [5] [4] [3] [2] ]1] [ENTER].3. Remove any load on the scale.4. Press [ENTER] to establish zero.5. Place a test weight on scale.6. Press [ZERO].7. Remove any load on the scale.8. Press [ENTER] to establish zero.9. Pause for motion delay.10. Place a test weight on scale.11. Enter the test weight value with the numeric keys.12. Press [ENTER] to establish span.13. Pause for motion delay.14. Press [ENTER] to accept calibration.15. Press [ENTER] to save calibration.16. Press [ENTER] to exit calibration.17. Remove the calibration weight.5.2.7 Multi-Point LinearizationIf the load cell signal input to the indicator has good repeatability and stability, thenusing multi-point linearization during calibration may significantly improve the ultimateaccuracy of the data displayed by the indicator. Parameter 119 (P119) enables ordisables this feature. Initially, the same basic procedures as a normal calibration areused to perform a multi-point linearization. All of the prompts provided will be exactlythe same as a standard calibration for each cal point up to the Cal ~ Good? prompt.If items are consistantly being weighed on either the high or low end of the cell capacity,the points of calibration can be skewed to either end of the spectrum. Multi-pointlinearization can compensate for a cell that is non-linear. Refer to page 117 for theinstructions on how to enter the calibration mode.Model 350 I.S. and Model 355 I.S. Service Manual 123
- Page 73 and 74: The default format code for all par
- Page 75 and 76: 4.12.2 Model 355 I.S.To sample usin
- Page 77 and 78: 4.14 Remote Key ConfigurationThe Mo
- Page 79 and 80: 4.16 Setpoint SetupThe Model 350/35
- Page 81 and 82: Pre-act values are entered in terms
- Page 83 and 84: 4. Next the PctHi value will be dis
- Page 85 and 86: 3. Press [TARE]. Once the tare key
- Page 87 and 88: 4.19.1 Enter a TargetPressing [TARE
- Page 89 and 90: ParameterSettingPre-Act 2 = 236 15,
- Page 91 and 92: 4.20.3 Start DischargeActivation Me
- Page 93 and 94: 3a. If a value is desired, enter it
- Page 95 and 96: 4.22 Absolute CheckweighThis progra
- Page 97 and 98: 4.23 Target Deviation CheckweighThe
- Page 99 and 100: 4.24 Batch 2The batch 2 program is
- Page 101 and 102: 4.25 Independent Setpoint Operation
- Page 103 and 104: 4.25.3 ExampleSetup a continuous-cy
- Page 105 and 106: Bars ON Weight Range (Lower Toleran
- Page 107 and 108: ParameterSettingDisplayedNameDescri
- Page 109 and 110: 4.27.2 Analog Output ExampleAnalogP
- Page 111 and 112: 4.28.2 Setup Master IndicatorSet th
- Page 113 and 114: ParameterSettingDisplayedNameDescri
- Page 115 and 116: Custom Transmit :1. [.002] [ENTER]
- Page 117 and 118: 4.28.5 300 Series Command CodesHex
- Page 119 and 120: 5 CalibrationThis chapter covers al
- Page 121 and 122: First Zero Calibration Method (355
- Page 123: 9. Pause for motion delay.10. Press
- Page 127 and 128: If the calibration weight value was
- Page 129 and 130: 6 Legal for TradeThe Model 350/355
- Page 131 and 132: 6.4.1 Physical SealThe most common
- Page 133 and 134: 7 TroubleshootingThis chapter conta
- Page 135 and 136: 7.1.4 Calibration ErrorsCodeF.S. ~T
- Page 137 and 138: Table 7.1 Information Parameters.Pa
- Page 139 and 140: 8 MaintenanceThe mainboard, keypad
- Page 141 and 142: 5. Connect the keypad cable to the
- Page 143 and 144: 9 Label DrawingsReference for Model
- Page 145 and 146: 9.5 355 I.S. Generic Rear Label (5
- Page 147 and 148: Appendix A: Model 350 I.S. and Mode
- Page 149 and 150: AWT35-500389ABModel 350 I.S. and Mo
- Page 151 and 152: AWT35-500389ABModel 350 I.S. and Mo
- Page 153 and 154: Uo or Voc or Vt ≤ Vmax, Io or Isc
- Page 156: Avery Weigh-Tronix USA1000 Armstron
13. Pause for motion delay.14. Press [TARE] to accept calibration.15. Press [TARE] to save calibration.16. Press [TARE] to exit calibration.17. Remove the calibration weight.Reset Calibration Gain Factors (355 I.S.):1. Press [1] [0] [0] [SELECT].2. Press [5] [4] [3] [2] ]1] [ENTER].3. Remove any load on the scale.4. Press [ENTER] to establish zero.5. Place a test weight on scale.6. Press [ZERO].7. Remove any load on the scale.8. Press [ENTER] to establish zero.9. Pause for motion delay.10. Place a test weight on scale.11. Enter the test weight value with the numeric keys.12. Press [ENTER] to establish span.13. Pause for motion delay.14. Press [ENTER] to accept calibration.15. Press [ENTER] to save calibration.16. Press [ENTER] to exit calibration.17. Remove the calibration weight.5.2.7 Multi-Point LinearizationIf the load cell signal input to the indicator has good repeatability and stability, thenusing multi-point linearization during calibration may significantly improve the ultimateaccuracy of the data displayed by the indicator. Parameter 119 (P119) enables ordisables this feature. Initially, the same basic procedures as a normal calibration areused to perform a multi-point linearization. All of the prompts provided will be exactlythe same as a standard calibration for each cal point up to the Cal ~ Good? prompt.If items are consistantly being weighed on either the high or low end of the cell capacity,the points of calibration can be skewed to either end of the spectrum. Multi-pointlinearization can compensate for a cell that is non-linear. Refer to page 117 for theinstructions on how to enter the calibration mode.Model 350 I.S. and Model 355 I.S. <strong>Service</strong> <strong>Manual</strong> 123